Lobanov, Y., Tong, C., Blundell, R., & Gol'tsman, G. (2009). A study of direct detection effect on the linearity of hot electron bolometer mixers. In Proc. 20th Int. Symp. Space Terahertz Technol. (pp. 282–287).
Abstract: We have performed a study of how direct detection affects the linearity and hence the calibration of an HEB mixer. Two types of waveguide HEB devices have been used: a 0.8 THz HEB mixer and a 1.0 THz HEB mixer which is ~5 times smaller than the former. Two independent experimental approaches were used. In the ΔG/G method, the conversion gain of the HEB mixer is first measured as a function of the bias current for a number of bias voltages. At each bias setting, we carefully measure the change in the operating current when the input loads are switched. From the measured data, we can derive the expected difference in gain between the hot and cold loads. In the second method (injection method [1]), the linearity of the HEB mixer is independently measured by injecting a modulated signal for different input load temperatures. The results of both approaches confirm that there is gain compression in the operation of HEB mixers. Based on the results of our measurements, we discuss the impact of direct detection effects on the operation of HEB mixers.
|
|
|
Lobanov, Y. V., Tong, C. - Y. E., Hedden, A. S., Blundell, R., & Gol'tsman, G. N. (2010). Microwave-assisted슠measurement슠of the슠frequency슠response슠of슠terahertz슠HEB슠mixers슠with a슠fourier슠transform슠spectrometer. In 21st International Symposium on Space Terahertz Technology (pp. 420–423).
Abstract: We describe a novel method of operation of the HEB direct detector for use with a Fourier Transform Spectrometer. Instead of elevating the bath temperature, we have measured the RF response of waveguide HEB mixers by applying microwave radiation to select appropriate bias conditions. In our experiment, a microwave signal is injected into the HEB mixer via its IF port. By choosing an appropriate injection level, the device can be operated close to the desired operating point. Furthermore, we have shown that both thermal biasing and microwave injection can reproduce the same spectral response of the HEB mixer. However, with the use of microwave injection, there is no need to wait for the mixer to reach thermal equilibrium, so characterisation can be done in less time. Also, the liquid helium consumption for our wet cryostat is also reduced. We have demonstrated that the signalto-noise ratio of the FTS measurements can be improved with microwave injection.
|
|
|
Trifonov, A., Tong, C. - Y. E., Grimes, P., Lobanov, Y., Kaurova, N., Blundell, R., et al. (2017). Development of A Silicon Membrane-based Multi-pixel Hot Electron Bolometer Receiver. In IEEE Trans. Appl. Supercond. (Vol. 27, 6).
Abstract: We report on the development of a multi-pixel
Hot Electron Bolometer (HEB) receiver fabricated using
silicon membrane technology. The receiver comprises a
2 × 2 array of four HEB mixers, fabricated on a single
chip. The HEB mixer chip is based on a superconducting
NbN thin film deposited on top of the silicon-on-insulator
(SOI) substrate. The thicknesses of the device layer and
handling layer of the SOI substrate are 20 μm and 300 μm
respectively. The thickness of the device layer is chosen
such that it corresponds to a quarter-wave in silicon at
1.35 THz. The HEB mixer is integrated with a bow-tie
antenna structure, in turn designed for coupling to a
circular waveguide,
|
|
|
Kovalyuk, V., Ferrari, S., Kahl, O., Semenov, A., Shcherbatenko, M., Lobanov, Y., et al. (2017). On-chip coherent detection with quantum limited sensitivity. Sci Rep, 7(1), 4812.
Abstract: While single photon detectors provide superior intensity sensitivity, spectral resolution is usually lost after the detection event. Yet for applications in low signal infrared spectroscopy recovering information about the photon's frequency contributions is essential. Here we use highly efficient waveguide integrated superconducting single-photon detectors for on-chip coherent detection. In a single nanophotonic device, we demonstrate both single-photon counting with up to 86% on-chip detection efficiency, as well as heterodyne coherent detection with spectral resolution f/f exceeding 10(11). By mixing a local oscillator with the single photon signal field, we observe frequency modulation at the intermediate frequency with ultra-low local oscillator power in the femto-Watt range. By optimizing the nanowire geometry and the working parameters of the detection scheme, we reach quantum-limited sensitivity. Our approach enables to realize matrix integrated heterodyne nanophotonic devices in the C-band wavelength range, for classical and quantum optics applications where single-photon counting as well as high spectral resolution are required simultaneously.
|
|
|
Kovalyuk, V., Ferrari, S., Kahl, O., Semenov, A., Lobanov, Y., Shcherbatenko, M., et al. (2017). Waveguide integrated superconducting single-photon detector for on-chip quantum and spectral photonic application. In J. Phys.: Conf. Ser. (Vol. 917, 062032).
Abstract: With use of the travelling-wave geometry approach, integrated superconductor- nanophotonic devices based on silicon nitride nanophotonic waveguide with a superconducting NbN-nanowire suited on top of the waveguide were fabricated. NbN-nanowire was operated as a single-photon counting detector with up to 92 % on-chip detection efficiency in the coherent mode, serving as a highly sensitive IR heterodyne mixer with spectral resolution (f/df) greater than 106 in C-band at 1550 nm wavelength
|
|