Lobanov, Y., Tong, C., Blundell, R., & Gol'tsman, G. (2009). A study of direct detection effect on the linearity of hot electron bolometer mixers. In Proc. 20th Int. Symp. Space Terahertz Technol. (pp. 282–287).
Abstract: We have performed a study of how direct detection affects the linearity and hence the calibration of an HEB mixer. Two types of waveguide HEB devices have been used: a 0.8 THz HEB mixer and a 1.0 THz HEB mixer which is ~5 times smaller than the former. Two independent experimental approaches were used. In the ΔG/G method, the conversion gain of the HEB mixer is first measured as a function of the bias current for a number of bias voltages. At each bias setting, we carefully measure the change in the operating current when the input loads are switched. From the measured data, we can derive the expected difference in gain between the hot and cold loads. In the second method (injection method [1]), the linearity of the HEB mixer is independently measured by injecting a modulated signal for different input load temperatures. The results of both approaches confirm that there is gain compression in the operation of HEB mixers. Based on the results of our measurements, we discuss the impact of direct detection effects on the operation of HEB mixers.
|
Kawamura, J., Blundell, R., Tong, C. - Y. E., Golts'man, G., Gershenzon, E., & Voronov B. (1996). Superconductive NbN hot-electron bolometric mixer performance at 250 GHz. In Proc. 7th Int. Symp. Space Terahertz Technol. (pp. 331–336).
Abstract: Thin film NbN (<40 A) strips are used as waveguide mixer elements. The electron cooling mechanism for the geometry is the electron-phonon interaction. We report a receiver noise temperature of 750 K at 244 GHz, with / IF = 1.5 GHz, Af= 500 MHz, and Tphysical = 4 K. The instantaneous bandwidth for this mixer is 1.6 GHz. The local oscillator (LO) power is 0.5 1.tW with 3 dB-uncertainty. The mixer is linear to 1 dB up to an input power level 6 dB below the LO power. We report the first detection of a molecular line emission using this class of mixer, and that the receiver noise temperature determined from Y-factor measurements reflects the true heterodyne sensitivity.
|
Loudkov, D., Tong, C. - Y. E., Marrone, D. P., Ryabchun, S., Paine, S. N., & Blundell, R. (2005). Transmission measurements of infrared filters for low-noise terahertz receiver applications. In Proc. 16th Int. Symp. Space Terahertz Technol. (pp. 354–357).
Abstract: Infrared (IR) filters are very important to the efficient operation of cryogenic receivers. Usually, such filters are mounted on the radiation shield of the cryostat to reduce the heat load to the 4 K stage. Insufficient filtering may cause the temperature of the mixing element in a receiver to be excessively warm, leading to degradation in sensitivity. These filters should be effective in blocking the room temperature IR radiation from outside the cryostat, yet should be transparent across the desired signal frequency band. In the Terahertz frequency range, which is close to the infrared, it is difficult to find an inexpensive low- loss material that can provide the required IR blocking capacity. We present transmission measurements, made using a Fourier Transform Spectrometer (FTS), of a number of potential infrared filters between 0.4 and 1.6 THz. The filters tested include the widely-used, Teflon-based, Zitex-A and Zitex-G films, alkali halide based infrared filter, and crystalline quartz coated with Parylene, and polyethylene films.
|
Tong, C. E., Blundell, R., Papa, D. C., Smith, M., Kawamura, J., Gol'tsman, G., et al. (1999). An all solid-state superconducting heterodyne receiver at terahertz frequencies. IEEE Microw. Guid. Wave Lett., 9(9), 366–368.
Abstract: A superconducting hot-electron bolometer mixer-receiver operating from 1 to 1.26 THz has been developed. This heterodyne receiver employs two solid-state local oscillators each consisting of a Gunn oscillator followed by two stages of varactor frequency multiplication. The measured receiver noise temperature is 1350 K at 1.035 THz and 2700 K at 1.26 THz. This receiver demonstrates that tunable solid-state local oscillators, supplying only a few micro-watts of output power, can be used in terahertz receiver applications.
|
Ryabchun, S., Tong, C. - Y. E., Blundell, R., Kimberk, R., & Gol'tsman, G. (2007). Study of the effect of microwave radiation on the operation of HEB mixers in the terahertz frequency range. IEEE Trans. Appl. Supercond., 17(2), 391–394.
Abstract: We have investigated the effect of injecting microwave radiation, with a frequency much lower than that corresponding to the energy gap of the superconductor, on the performance of the hot-electron bolometer mixer incorporated into a THz heterodyne receiver. More specifically, we show that exposing the mixer to microwave radiation does not cause a significant rise of the receiver noise temperature and fall of the mixer conversion gain so long as the microwave power is a small fraction of local oscillator power. The injection of a small, but controlled amount of microwave power therefore enables active compensation of local oscillator power and coupling fluctuations which can significantly degrade the gain stability of hot electron bolometer mixer receivers.
|