Ryabchun, S., Tong, C. -yu E., Blundell, R., Kimberk, R., & Gol’tsman, G. (2007). Stabilisation of a terahertz hot-electron bolometer mixer with microwave feedback control. In Proc. 18th Int. Symp. Space Terahertz Technol. (pp. 193–198).
Abstract: We report on implementation of microwave feedback control loop to stabilise the performance of an HEB mixer receiver. It is shown that the receiver sensitivity increases by a factor of 4 over a 16-minute scan, and the corresponding Allan time increases up to 10 seconds, as opposed to an open loop value of 1 second. Our experiments also demonstrate that the receiver sensitivity is limited by the intermediate frequency chain.
|
|
|
Heusinger, M. A., Nebosis, R. S., Schatz, W., Renk, K. F., Gol’tsman, G. N., Karasik, B. S., et al. (1993). Temperature dependence of bolometric and non-bolometric photoresponse of a structured YBa2Cu3O7-δ thin film. In M. Meissner, & R. O. Pohl (Eds.), Phonon Scattering in Condensed Matter VII. Springer Series in Solid-State Sciences (Vol. 112, pp. 193–195).
Abstract: We investigated the temperature dependence of the transient voltage photoresponse of a current biased structured YBa2Cu3O7−δ thin film in its transition temperature region, around 79 K. Both, picosecond nonbolometric and nanosecond bolometric response to ultrashort far-infrared laser pulses were found for frequencies between 25 cm−1 and 215 cm−1. We will discuss optimum conditions for radiation detection and present an analysis of the dynamical behaviour of excited high T c thin films.
|
|
|
Gershenzon, E. M., Goltsman, G. N., Semenov, A. D., & Sergeev, A. V. (1989). Limiting characteristic of fast superconducting bolometers. Sov. Phys.-Tech. Phys., 34, 195–199.
Abstract: Теоретически и экспериментально исследовано физическое ограничение быстродействия сверхпроводящего болометра. Показано, что минимальная постоянная времени реализуется в условиях электронного разогрева и определяется процессом неупругого электрон-фонон- ного взаимодействия. Сформулированы требования кконструкции «электронного болометра» для достижения предельной чувствительности. Проведено сравнение характеристик электронного болометра и обычных болометров различных типов.
|
|
|
Huebers, H. - W., Semenov, A., Schubert, J., Gol’tsman, G. N., Voronov, B. M., Gershenzon, E. M., et al. (2000). NbN hot-electron bolometer as THz mixer for SOFIA. In R. K. Melugin, & H. - P. Roeser (Eds.), Proc. SPIE (Vol. 4014, pp. 195–202). SPIE.
Abstract: Heterodyne receivers for applications in astronomy need quantum limited sensitivity. We have investigated phonon- cooled NbN hot electron bolometric mixers in the frequency range from 0.7 THz to 5.2 THz. The devices were 3.5 nm thin films with an in-plane dimension of 1.7 X 0.2 micrometers 2 integrated in a complementary logarithmic spiral antenna. The best measured DSB receiver noise temperatures are 1300 K (0.7 THz), 2000 K (1.4 THz), 2100 K (1.6 THz), 2600 K (2.5 THz), 4000 K (3.1 THz), 5600 K (4.3 THz), and 8800 K (5.2 THz). The sensitivity fluctuation, the long term stability, and the antenna pattern were measured. The results demonstrate that this mixer is very well suited for GREAT, the German heterodyne receiver for SOFIA.
|
|
|
Gol'tsman, G. N. (1999). Hot electron bolometric mixers: new terahertz technology. Infrared Physics & Technology, 40(3), 199–206.
Abstract: This paper presents an overview of recent results for NbN phonon-cooled hot electron bolometric (HEB) mixers. The noise temperature of the receivers based on both quasioptical and waveguide versions of HEB mixers has crossed the level of 1 K GHz−1 at 430 GHz (410 K), 600–650 GHz (480 K), 750 GHz (600 K), 810 GHz (780 K) and is close to that level at 1.1 THz (1250 K) and 2.5 THz (4500 K). The gain bandwidth measured for quasioptical HEB mixer at 620 GHz reached 4 GHz and the noise temperature bandwidth was almost 8 GHz. Local oscillator power requirements are about 1 μW for mixers made by photolithography and about 100 nW for mixers made by e-beam lithography. A waveguide version of 800 GHz receiver was installed at the Submillimeter Telescope Observatory on Mt. Graham, AZ, to conduct astronomical observations of known submillimeter lines (CO, J=7→6, CI, J=2→1). It was proved that the receiver works as a practical instrument.
|
|