Gol'tsman, G. N., Elant'iev, A. I., Karasik, B. S., & Gershenzon, E. M. (1993). Antenna – coupled superconducting electron-heating bolometer. In Proc. 4th Int. Symp. Space Terahertz Technol. (pp. 623–628).
Abstract: We propose a novel antenna-coupled superconducting bolometer based on electron-heating in the resistive state. A short narrow ultrathin super- conducting film strip (sized approximately 4x1x0.01 pm 3 ), which is in good thermal contact with the thermostat, serves as a resistive load for infrared or submillimeter current. In contrast to conventional isothermal super- conducting bolometers electron-heating ones can have a higher sensitivity which grows when filni. thickness is reduced. Response time of electron- heating bolometer does not depend on heat transfer from the film to the enviroment. To calculate the sensitivity (NEP), we have used experimental data on wideband Al, Nb and NbN bolometers which have the same un- derlying physical mechanism. The bolom.eters have been made in the form of a structure composed of a number of long narrow strips. The values of for Al, NEP have been found to be 1.5 . 113 -16 1 140 -15 ) and 2 . 10 – 14werT,-1/2 – Nb and NbN respectively. In the paper, the prospects are also discussed of improving the picosecond YBaCuO detector, developed recently. NEP value of the detector, if combined with a microantenna, can reach the order of 10- •ilz-v2.
|
Gershenzon, E. M., Gershenson, M. E., Goltsman, G. N., Lyulkin, A. M., Semenov, A. D., & Sergeev, A. V. (1989). Limiting characteristics of fast-response superconducting bolometers. Zhurnal Tekhnicheskoi Fiziki, 59(2), 11–120.
Abstract: Теоретически и экспериментально исследовано физическое ограничение быстродействия сверхпроводящего болометра. Показано, что минимальная постоянная времени реализуется в условиях электронного разогрева и определяется процессом неупругого электрон-фонон-ного взаимодействия. Сформулированы требования к конструкции «электронного болометра» для достижения предельной чувствительности. Проведено сравнение характеристик электронного болометра и обычных болометров различных типов.
|
Gershenzon, E. M., Gershenzon, M. E., Gol'tsman, G. N., Semenov, A. D., & Sergeev, A. V. (1982). Nonselective effect of electromagnetic radiation on a superconducting film in the resistive state. JETP Lett., 36(7), 296–299.
|
Kitaeva, G. K., Kornienko, V. V., Kuznetsov, K. A., Pentin, I. V., Smirnov, K. V., & Vakhtomin, Y. B. (2019). Direct detection of the idler THz radiation generated by spontaneous parametric down-conversion. Opt. Lett., 44(5), 1198–1201.
Abstract: We study parametric down-conversion (PDC) of optical laser radiation in the strongly frequency non-degenerate regime which is promising for the generation of quantum-correlated pairs of extremely different spectral ranges, the optical and the terahertz (THz) ones. The possibility to detect tenuous THz-frequency photon fluxes generated under low-gain spontaneous PDC is demonstrated using a hot electron bolometer. Then experimental dependences of the THz radiation power on the detection angle and on the pump intensity are analyzed.
|
Ozhegov, R. V., Gorshkov, K. N., Okunev, O. V., & Gol’tsman, G. N. (2010). Superconducting hot-electron bolometer mixer as element of thermal imager matrix. Tech. Phys. Lett., 36(11), 1006–1008.
Abstract: The possibility of using a matrix of sensitive elements on a 12-mm-diameter hyperhemispherical lens in a thermal imager operating in the terahertz range has been studied. Dimensions of a lens region acceptable for arrangement of the matrix, in which the receiver noise temperature varies within 16% of the mean value, are determined to be 3.3% of the lens diameter. Deviations of the main lobe of the directivity pattern are evaluated, which amount to ±1.25° relative to the direction toward the optimum position of a mixer. The fluctuation sensitivity of the receiver measured in experiment is 0.5 K at a frequency of 300 GHz.
|