Kawamura, J., Tong, C. - Y. E., Blundell, R., Papa, D. C., Hunter, T. R., Patt, F., et al. (2001). Terahertz-frequency waveguide NbN hot-electron bolometer mixer. IEEE Trans. Appl. Supercond., 11(1), 952–954.
Abstract: We have developed a low-noise waveguide heterodyne receiver for operation near 1 THz using phonon-cooled NbN hot-electron bolometers. The mixer elements are submicron-sized microbridges of 4 nm-thick NbN film fabricated on a quartz substrate. Operating at a bath temperature of 4.2 K, the double-sideband receiver noise temperature is 760 K at 1.02 THz and 1100 K at 1.26 THz. The local oscillator is provided by solid-state sources, and power measured at the source is less than 1 /spl mu/W. The intermediate frequency bandwidth exceeds 2 GHz. The receiver was used to make the first ground-based heterodyne detection of a celestial spectroscopic line above 1 THz.
|
|
|
Antipov, S. V., Svechnikov, S. I., Smirnov, K. V., Vakhtomin, Y. B., Finkel, M. I., Goltsman, G. N., et al. (2001). Noise temperature of quasioptical NbN hot electron bolometer mixers at 900 GHz. Physics of Vibrations, 9(4), 242–245.
|
|
|
Svechnikov, S. I., Antipov, S. V., Vakhtomin, Y. B., Goltsman, G. N., Gershenzon, E. M., Cherednichenko, S. I., et al. (2001). Conversion and noise bandwidths of terahertz NbN hot-electron bolometer mixers. Physics of Vibrations, 9(3), 205–210.
|
|
|
Hübers, H. - W., Semenov, A., Schubert, J., Gol'tsman, G., Voronov, B., & Gershenzon, E. (2000). Performance of the phonon-cooled hot-electron bolometric mixer between 0.7 THz and 5.2 THz. In Proc. 8-th Int. Conf. on Terahertz Electronics (pp. 117–119).
Abstract: We report on the phonon cooled NbN hot electron bolometer as mixer in the terahertz frequency range. Its hybrid antenna consists of a hyperhemispheric silicon lens and a logarithmic-spiral feed antenna. Noise temperatures have been measured between 0.7 THz and 5.2 THz. A quarter wavelength layer of Parylene works as antireflection coating for the silicon lens and reduces the noise temperature by about 30. It was found that the antenna pattern at 2.5 THz is determined by the feed antenna and not by the diameter of the lens.
|
|
|
Gerecht, E., Musante, C. F., Zhuang, Y., Yngvesson, K. S., Gol’tsman, G. N., Voronov, B. M., et al. (1999). NbN hot electron bolometric mixerss—a new technology for low-noise THz receivers. IEEE Trans. Appl. Supercond., 47(12), 2519–2527.
Abstract: New advances in hot electron bolometer (HEB) mixers have recently resulted in record-low receiver noise temperatures at terahertz frequencies. We have developed quasi-optically coupled NbN HEB mixers and measured noise temperatures up to 2.24 THz, as described in this paper. We project the anticipated future performance of such receivers to have even lower noise temperature and local-oscillator power requirement as well as wider gain and noise bandwidths. We introduce a proposal for integrated focal plane arrays of HEB mixers that will further increase the detection speed of terahertz systems.
|
|