Tong, C. - Y. E., Trifonov, A., Shurakov, A., Blundell, R., & Gol’tsman, G. (2015). A microwave-operated hot-electron-bolometric power detector for terahertz radiation. IEEE Trans. Appl. Supercond., 25(3), 2300604 (1 to 4).
Abstract: A new class of microwave-operated THz power detectors based on the NbN hot-electron-bolometer (HEB) mixer is proposed. The injected microwave signal ( 1 GHz) serves the dual purpose of pumping the HEB element and enabling the read-out of the internal state of the device. A cryogenic amplifier amplifies the reflected microwave signal from the device and a homodyne scheme recovers the effects of the incident THz radiation. Two modes of operation have been identified, depending on the level of incident radiation. For weak signals, we use a chopper to chop the incident radiation against a black body reference and a lock-in amplifier to perform synchronous detection of the homodyne readout. The voltage measured is proportional to the incident power, and we estimate an optical noise equivalent power of 5pW/ √Hz at 0.83 THz. At higher signal levels, the homodyne circuit recovers the stream of steady relaxation oscillation pulses from the HEB device. The frequency of these pulses is in the MHz frequency range and bears a linear relationship with the incident THz radiation over an input power range of 15 dB. A digital frequency counter is used to measure THz power. The applicable power range is between 1 nW and 1 μW.
|
|
|
Shurakov, A., Tong, C. - Y. E., Blundell, R., Kaurova, N., Voronov, B., & Gol'tsman, G. (2013). Microwave stabilization of a HEB mixer in a pulse-tube cryocooler. IEEE Trans. Appl. Supercond., 23(3), 1501504.
Abstract: We report the results of our study of the stability of an 800 GHz hot electron bolometer (HEB) mixer cooled with a pulse-tube cryocooler. Pulse-tube cryocoolers introduce temperature fluctuations as well as mechanical vibrations at a frequency of ~1 Hz, both of which can cause receiver gain fluctuations at that frequency. In our system, the motor of the cryocooler was separated from the cryostat to minimize mechanical vibrations, leaving thermal effects as the dominant source of the receiver gain fluctuations. We measured root mean square temperature variations of the 4 K stage of ~7 mK. The HEB mixer was pumped by a solid state local oscillator at 810 GHz. The root mean square current fluctuations at the low noise operating point (1.50 mV, 56.5 μA) were ~0.12 μA, and were predominantly due to thermal fluctuations. To stabilize the bias current, microwave radiation was injected to the HEB mixer. The injected power level was set by a proportional-integral-derivative controller, which completely compensates for the bias current oscillations induced by the pulse-tube cryocooler. Significant improvement in the Allan variance of the receiver output power was obtained, and an Allan time of 5 s was measured.
|
|
|
Shurakov, A., Tong, C. -yu E., Grimes, P., Blundell, R., & Golt'sman, G. (2015). A microwave reflection readout scheme for hot electron bolometric direct detector. IEEE Trans. THz Sci. Technol., 5, 81–84.
Abstract: In this paper, we propose and present data from a fast THz detector based on the repurpose of hot electron bolometer mixers (HEB) fabricated from superconducting NbN thinfilm. This detector is essentially a traditional NbN bolometer element that operates under the influence of a microwave pump. The in-jected microwave power serves the dual purpose of enhancing the detector sensitivity and reading out the impedance changes of the device in response to incidentTHz radiation. We have measured an optical Noise Equivalent Power of 4 pW/ Hz for our detector at a bath temperature of 4.2 K. The measurement frequency was 0.83 THz and the modulation frequency was 1.48 kHz. The readout
scheme is versatile and facilitates both high-speed operation as well as multi-pixel applications.
|
|
|
Shurakov, A., Seliverstov, S., Kaurova, N., Finkel, M., Voronov, B., & Goltsman, G. (2012). Input bandwidth of hot electron bolometer with spiral antenna. IEEE Trans. THz Sci. Technol., 2(4), 400–405.
Abstract: We report the results of our study of the input bandwidth of hot electron bolometers (HEB) embedded into the planar log-spiral antenna. The sensitive element is made of the ultrathin superconducting NbN film patterned as a bridge at the feed of the antenna. The contacts between the antenna and a sensitive element are made from in situ deposited gold (i.e., deposited over NbN film without breaking vacuum), which gives high quality contacts and makes the response of the HEB at higher frequencies less affected by the RF loss. An accurate experimental spectroscopic procedure is demonstrated that leads to the confirmation of the wide ( 8 THz) bandwidth in this antenna coupled device.
|
|
|
Tretyakov, I., Svyatodukh, S., Chumakova, A., Perepelitsa, A., Kaurova, N., Shurakov, A., et al. (2019). Room temperature silicon detector for IR range coated with Ag2S quantum dots. In IRMMW-THz.
Abstract: A silicon has been the chief technological semiconducting material of modern microelectronics and has had a strong influence on all aspects of society. Applications of Si-based optoelectronic devices are limited to the visible and near infrared ranges. The expansion of the Si absorption to shorter wavelengths of the infrared range is of considerable interest to optoelectronic applications. By creating impurity states in Si it is possible to cause sub-band gap photon absorption. Here, we present an elegant and effective technology of extending the photoresponse of towards the IR range. Our approach is based on the use of Ag 2 S quantum dots (QDs) planted on the surface of Si. The specific sensitivity of the Ag 2 S/Si heterostructure is 10 11 cm√HzW -1 at 1.55μm. Our findings open a path towards the future study and development of Si detectors for technological applications.
|
|