Miao, W., Zhang, W., Zhong, J. Q., Shi, S. C., Delorme, Y., Lefevre, R., et al. (2014). Non-uniform absorption of terahertz radiation on superconducting hot electron bolometer microbridges. <ef><bf><bc>Appl. Phys. Lett., 104, 052605(1–4).
Abstract: We interpret the experimental observation of a frequency-dependence of superconducting hot electron bolometer (HEB) mixers by taking into account the non-uniform absorption of the terahertz radiation on the superconducting HEB microbridge. The radiation absorption is assumed to be proportional to the local surface resistance of the HEB microbridge, which is computed using the Mattis-Bardeen theory. With this assumption the dc and mixing characteristics of a superconducting niobium-nitride (NbN) HEB device have been modeled at frequencies below and above the equilibrium gap frequency of the NbN film.
|
Zhang, W., Miao, W., Zhong, J. Q., Shi, S. C., Hayton, D. J., Vercruyssen, N., et al. (2014). Temperature dependence of the receiver noise temperature and IF bandwidth of superconducting hot electron bolometer mixers. Supercond. Sci. Technol., 27(8), 085013 (1 to 5).
Abstract: In this paper we study the temperature dependence of the receiver noise temperature and IF noise bandwidth of superconducting hot electron bolometer (HEB) mixers. Three superconducting NbN HEB devices of different transition temperatures (Tc) are measured at 0.85 THz and 1.4 THz at different bath temperatures (Tbath) between 4 K and 9 K. Measurement results demonstrate that the receiver noise temperature of superconducting NbN HEB devices is nearly constant for Tbath/Tc, less than 0.8, which is consistent with the simulation based on a distributed hot-spot model. In addition, the IF noise bandwidth appears independent of Tbath/Tc, indicating the dominance of phonon cooling in the investigated HEB devices.
|
Zhang, W., Miao, W., Yao, Q. J., Lin, Z. H., Shi, S. C., Gao, J. R., et al. (2012). Spectral response and noise temperature of a 2.5 THz spiral antenna coupled NbN HEB mixer. Phys. Procedia, 36, 334–337.
Abstract: We report on a 2.5 THz spiral antenna coupled NbN hot electron bolometer (HEB) mixers, fabricated with in-situ process. The receiver noise temperature with lowest value of 1180 K is in good agreement with calculated quantum efficiency factor as a function of bias voltage. In addition, the measured spectral response of the spiral antenna coupled NbN HEB mixer shows broad frequency coverage of 0.8-3 THz, and corrected response for optical losses, FTS, and coupling efficiency between antenna and bolometer falls with frequency due to diffraction-limited beam of lens/antenna combination.
|
Zhang, W., Miao, W., Zhong, J. Q., Shi, S. C., Hayton, D. J., Vercruyssen, N., et al. (2013). Temperature dependence of superconducting hot electron bolometers. In Not published results: 24th international symposium on space terahertz technology.
|
Ren, Y., Zhang, D. X., Zhou, K. M., Miao, W., Zhang, W., Shi, S. C., et al. (2019). 10.6 μm heterodyne receiver based on a superconducting hot-electron bolometer mixer and a quantum cascade laser. AIP Advances, 9(7), 075307.
Abstract: We report on the development of a heterodyne receiver at mid-infrared wavelength for high-resolution spectroscopy applications. The receiver employs a superconducting NbN hot electron bolometer as a mixer and a room temperature distributed feedback quantum cascade laser operating at 10.6 μm (28.2 THz) as a local oscillator. The stabilization of the heterodyne receiver has been achieved using a feedback loop controlling the output power of the laser. Improved Allan variance times as well as a double sideband receiver noise temperature of 5000 K and a noise bandwidth of 2.8 GHz of the receiver system are demonstrated.
The work is supported in part by the National Key R&D Program of China under Grant 2018YFA0404701, by the CAS program under Grant QYZDJ-SSW-SLH043 and GJJSTD20180003, by the National Natural Science Foundation of China (NSFC) under Grant 11773083, by the “Hundred Talents Program” of the “Pioneer Initiative”, and in part by the CAS Key Lab for Radio Astronomy.
|