Zhang, J., Pearlman, A., Slysz, W., Verevkin, A., Sobolewski, R., Wilsher, K., et al. (2003). A superconducting single-photon detector for CMOS IC probing. In Proc. 16-th LEOS (Vol. 2, pp. 602–603).
Abstract: In this paper, a novel, time-resolved, NbN-based, superconducting single-photon detector (SSPD) has been developed for probing CMOS integrated circuits (ICs) using photon emission timing analysis (PETA).
|
|
|
Tong, C. - Y. E., Trifonov, A., Shurakov, A., Blundell, R., & Gol’tsman, G. (2015). A microwave-operated hot-electron-bolometric power detector for terahertz radiation. IEEE Trans. Appl. Supercond., 25(3), 2300604 (1 to 4).
Abstract: A new class of microwave-operated THz power detectors based on the NbN hot-electron-bolometer (HEB) mixer is proposed. The injected microwave signal ( 1 GHz) serves the dual purpose of pumping the HEB element and enabling the read-out of the internal state of the device. A cryogenic amplifier amplifies the reflected microwave signal from the device and a homodyne scheme recovers the effects of the incident THz radiation. Two modes of operation have been identified, depending on the level of incident radiation. For weak signals, we use a chopper to chop the incident radiation against a black body reference and a lock-in amplifier to perform synchronous detection of the homodyne readout. The voltage measured is proportional to the incident power, and we estimate an optical noise equivalent power of 5pW/ √Hz at 0.83 THz. At higher signal levels, the homodyne circuit recovers the stream of steady relaxation oscillation pulses from the HEB device. The frequency of these pulses is in the MHz frequency range and bears a linear relationship with the incident THz radiation over an input power range of 15 dB. A digital frequency counter is used to measure THz power. The applicable power range is between 1 nW and 1 μW.
|
|
|
Blundell, R., Kawamura, J. H., Tong, C. E., Papa, D. C., Hunter, T. R., Gol’tsman, G. N., et al. (1998). A hot-electron bolometer mixer receiver for the 680-830 GHz frequency range. In Proc. 6-th Int. Conf. Terahertz Electron. (pp. 18–20). IEEE.
Abstract: We describe a heterodyne receiver designed to operate in the partially transparent atmospheric windows centered on 680 and 830 GHz. The receiver incorporates a niobium nitride thin film, cooled to 4.2 K, as the phonon-cooled hot-electron mixer element. The double sideband receiver noise, measured over the frequency range 680-830 GHz, is typically 700-1300 K. The instantaneous output bandwidth of the receiver is 600 MHz. This receiver has recently been used at the SubMillimeter Telescope, jointly operated by the Steward Observatory and the Max Planck Institute for Radioastronomy, for observations of the neutral carbon and CO spectral lines at 810 GHz and at 806 and 691 GHz respectively. Laboratory measurements on a second mixer in the same test receiver have yielded extended high frequency performance to 1 THz.
|
|
|
Ejrnaes, M., Cristiano, R., Quaranta, O., Pagano, S., Gaggero, A., Mattioli, F., et al. (2007). A cascade switching superconducting single photon detector. Appl. Phys. Lett., 91(26), 262509 (1 to 3).
Abstract: We have realized superconducting single photon detectors with reduced inductance and increased signal pulse amplitude. The detectors are based on a parallel connection of ultrathin NbN nanowires with a common bias inductance. When properly biased, an absorbed photon induces a cascade switch of all the parallel wires generating a signal pulse amplitude of 2mV. The parallel wire configuration lowers the detector inductance and reduces the response time well below 1ns.
This work was performed in the framework of the EU project “SINPHONIA” NMP4-CT-2005-016433.
|
|
|
Tong, C. - Y. E., Meledin, D., Loudkov, D., Blundell, R., Erickson, N., Kawamura, J., et al. (2003). A 1.5 THz Hot-Electron Bolometer mixer operated by a planar diode based local oscillator. In IEEE MTT-S Int. Microwave Symp. Digest (Vol. 2, pp. 751–754).
Abstract: We have developed a 1.5 THz superconducting NbN Hot-Electron Bolometer mixer. It is operated by an all-solid-state Local Oscillator comprising of a cascade of 4 planar doublers following an MMIC based W-band power amplifier. The threshold available pump power is estimated to be 1 /spl mu/W.
|
|