Baselmans, J. J. A., Hajenius, M., Gao, J. R., Baryshev, A., Kooi, J., Klapwijk, T. M., et al. (2004). Hot electron bolometer mixers with improved interfaces: sensitivity, LO power and stability. In Proc. 15th Int. Symp. Space Terahertz Technol. (pp. 17–24).
Abstract: We study twin slot antenna coupled NbN hot electron bolometer mixers with an improved contact structure and a small volume, ranging from 1 µm × 0.1 µm to 2 × 0.3 µm. We obtain a DSB receiver noise temperature of 900 K at 1.6 THz and 940 K at 1.9 THz. To explore the practical usability of such small HEB mixers we evaluate the LO power requirement, the sensitivity and the stability. We find that the LO power requirement of the smallest mixers is reduced to about 240 nW at the Si lens of the mixer. This value is larger than expected from the isothermal technique and the known losses in the lens by a factor of 3-3.5. The stability of these receivers is characterized using a measurement of the Allan Variance. We find an Allan time of 0.5 sec. in an 80 MHz bandwidth. A small increase in stability can be reached by using a higher bias at the expense of a significant amount of sensitivity. The stability is sufficient for spectroscopic applications in a 1 MHz bandwidth at a 1 Hz chopping frequency.
|
Hajenius, M., Baselmans, J. J. A., Gao, J. R., Klapwijk, T. M., de Korte 2, P. A. J., Voronov, B., et al. (2004). Increased bandwidth of NbN phonon cooled hot electron bolometer mixers. In Proc. 15th Int. Symp. Space Terahertz Technol. (pp. 381–386).
Abstract: We study experimentally the IF gain bandwidth of NbN phonon-cooled hot-electron-bolometer (HEB) mixers for a set of devices with different contact structures but an identical NbN film. We observe that the IF bandwidth depends strongly on the exact contact structure and find an IF gain bandwidth of 6 GHz for a device with an additional superconducting layer (NbTiN) in between the active NbN film and the gold contact to the antenna. These results contradict the common opinion that the IF bandwidth is determined by the phonon-escape time between the NbN film and the substrate. Hence we calculate the IF gain bandwidth of a superconducting film using a two-temperature model. We find that the bandwidth increases strongly with operating temperature and is not limited by the phonon escape time. This is because of strong temperature dependence of the phonon specific heat in the NbN film.
|
Gao, J. R., Hajenius, M., Baselmans, J. J. A., Klapwijk, T. M., de Korte, P. A. J., Voronov, B., et al. (2004). NbN hot electron bolometer mixers with superior performance for space applications. In E. Armandillo, & B. Leone (Eds.), Proc. Int. workshop on low temp. electronics (pp. 11–17). Noordwijk.
|
Baselmans, J. J. A., Hajenius, M., Gao, J. R., Klapwijk, T. M., de Korte, P. A. J., Voronov, B., et al. (2003). Noise performance of NbN hot electron bolometer mixers at 2.5 THz and its dependence on the contact resistance. In Proc. 14th Int. Symp. Space Terahertz Technol. (pp. 11–19).
Abstract: NbN hot electron bolometer mixers (HEBM) are at this moment the best heterodyne receivers for frequencies above 1 Thz. However, the fabrication procedure of these devices is such that the quality of the interface between the NbN superconducting film and the contact structure is not under good control. The result is a low transparency interface between the bolometer itself and the contact/antenna structure. In this paper we report a detailed experimental study on a novel idea to increase the transparency of this interface. This leads to a record sensitivity and more reproducible performance. We compare identical bolometers, coupled with a spiral antenna, with different NbN bolometer-contact pad interfaces. We find that cleaning the NbN interface alone results in an increase in the noise temperature. However, cleaning the NbN interface and adding a thin additional superconductor prior to the gold contact deposition improves the noise temperature of the HEBm with more than a factor of 2. A device with a contact pad on top of an in-situ cleaned NbN film consisting of 10 nm of NbTiN and 40 nm of gold has a DSB noise temperature of 1050 K at 2.5 THz.
|
Baselmans, J. J. A., Hajenius, M., Gao, J., de Korte, P., Klapwijk, T. M., Voronov, B., et al. (2004). Doubling of sensitivity and bandwidth in phonon-cooled hot-electron bolometer mixers. In J. Zmuidzinas, W. S. Holland, & S. Withington (Eds.), Proc. SPIE (Vol. 5498, pp. 168–176). SPIE.
Abstract: NbN hot electron bolometer (HEB) mixers are at this moment the best heterodyne detectors for frequencies above 1 THz. However, the fabrication procedure of these devices is such that the quality of the interface between the NbN superconducting film and the contact structure is not under good control. This results in a contact resistance between the NbN bolometer and the contact pad. We compare identical bolometers, with different NbN – contact pad interfaces, coupled with a spiral antenna. We find that cleaning the NbN interface and adding a thin additional superconductor prior to the gold contact deposition improves the noise temperature and the bandwidth of the HEB mixers with more than a factor of 2. We obtain a DSB noise temperature of 950 K at 2.5 THz and a Gain bandwidth of 5-6 GHz. For use in real receiver systems we design small volume (0.15x1 micron) HEB mixers with a twin slot antenna. We find that these mixers combine good sensitivity (900 K at 1.6 THz) with low LO power requirement, which is 160 – 240 nW at the Si lens of the mixer. This value is larger than expected from the isothermal technique and the known losses in the lens by a factor of 3-3.5.
|
Hajenius, M., Baselmans, J. J. A., Gao, J. R., Klapwijk, T. M., de Korte, P. A. J., Voronov, B., et al. (2004). Low noise NbN superconducting hot electron bolometer mixers at 1.9 and 2.5 THz. Supercond. Sci. Technol., 17(5), S224–S228.
Abstract: NbN phonon-cooled hot electron bolometer mixers (HEBs) have been realized with negligible contact resistance between the bolometer itself and the contact structure. Using a combination of in situ cleaning of the NbN film and the use of an additional superconducting interlayer of a 10 nm NbTiN layer between the Au of the contact structure and the NbN film superior noise temperatures have been obtained as low as 950 K at 2.5 THz and 750 K at 1.9 THz. Here we address in detail the DC characterization of these devices, the interface transparencies between the bolometers and the contacts and the consequences of these factors on the mixer performance.
|
Baselmans, J. J. A., Baryshev, A., Reker, S. F., Hajenius, M., Gao, J. R., Klapwijk, T. M., et al. (2005). Direct detection effect in small volume hot electron bolometer mixers. Appl. Phys. Lett., 86(16), 163503 (1 to 3).
Abstract: We measure the direct detection effect in a small volume (0.15μm×1μm×3.5nm)(0.15μm×1μm×3.5nm) quasioptical NbN phonon cooled hot electronbolometermixer at 1.6THz1.6THz. We find that the small signal sensitivity of the receiver is underestimated by 35% due to the direct detection effect and that the optimal operating point is shifted to higher bias voltages when using calibration loads of 300K300K and 77K77K. Using a 200GHz200GHzbandpass filter at 4.2K4.2K the direct detection effect virtually disappears. This has important implications for the calibration procedure of these receivers in real telescope systems.
|
Hajenius, M., Baselmans, J. J. A., Baryshev, A., Gao, J. R., Klapwijk, T. M., Kooi, J. W., et al. (2006). Full characterization and analysis of a terahertz heterodyne receiver based on a NbN hot electron bolometer. J. Appl. Phys., 100(7), 074507.
|
de Lange, G., Kuipers, J. J., Klapwijk, T. M., Panhuyzen, R. A., van de Stadt, H., & de Graauw, M. W. M. (1995). Superconducting resonator circuits at frequencies above the gap frequency. J. Appl. Phys., 77(4), 1795–1804.
|
Khosropanah, P., Gao, J. R., Laauwen, W. M., Hajenius, M., & Klapwijk, T. M. (2007). Low noise NbN hot electron bolometer mixer at 4.3 THz. Appl. Phys. Lett., 91, 221111 (1 to 3).
Abstract: We have studied the sensitivity of a superconducting NbN hot electron bolometer mixer integrated with a spiral antenna at 4.3 THz. Using hot/cold blackbody loads and a beam splitter all in vacuum, we measured a double sideband receiver noise temperature of 1300 K at the optimum local oscillator (LO) power of 330 nW, which is about 12 times the quantum noise (hnu/2kB). Our result indicates that there is no sign of degradation of the mixing process at the superterahertz frequencies. Moreover, a measurement method is introduced which allows us for an accurate determination of the sensitivity despite LO power fluctuations.
|
Ganzevles, W. F. M., Gao, J. R., de Korte, P. A. J., & Klapwijk, T. M. (2001). Direct response of microstrip line coupled Nb THz hot-electron bolometer mixers. Appl. Phys. Lett., 79(15), 2483–2485.
|
Jackson, B. D., Baryshev, A. M., de Lange, G., Gao, J. R., Shitov, S. V., Iosad, N. N., et al. (2001). Low-noise 1 THz superconductor-insulator-superconductor mixer incorporating a NbTiN/SiO2/Al tuning circuit. Appl. Phys. Lett., 79(3), 436.
|
Baselmans, J. J. A., Hajenius, M., Gao, J. R., Klapwijk, T. M., de Korte, P. A. J., Voronov, B., et al. (2004). Doubling of sensitivity and bandwidth in phonon cooled hot electron bolometer mixers. Appl. Phys. Lett., 84(11), 1958–1960.
Abstract: We demonstrate that the performance of NbN lattice cooled hot electron bolometer mixers depends strongly on the interface quality between the bolometer and the contact structure. We show experimentally that both the receiver noise temperature and the gain bandwidth can be improved by more than a factor of 2 by cleaning the interface and adding an additional superconducting interlayer to the contact pad. Using this we obtain a double sideband receiver noise temperature TN,DSB=950 K
at 2.5 THz and 4.3 K, uncorrected for losses in the optics. At the same bias point, we obtain an IF gain bandwidth of 6 GHz.
|
Dorenbos, S. N., Reiger, E. M., Perinetti, U., Zwiller, V., Zijlstra, T., & Klapwijk, T. M. (2008). Low noise superconducting single photon detectors on silicon. Appl. Phys. Lett., 93(13), 131101.
|
Floet, D. W., Baselmans, J. J. A., Klapwijk, T. M., & Gao, J. R. (1998). Resistive transition of niobium superconducting hot-electron bolometer mixers. Appl. Phys. Lett., 73(19), 2826.
|