Lobanov, Y., Tong, C., Blundell, R., & Gol'tsman, G. (2009). A study of direct detection effect on the linearity of hot electron bolometer mixers. In Proc. 20th Int. Symp. Space Terahertz Technol. (pp. 282–287).
Abstract: We have performed a study of how direct detection affects the linearity and hence the calibration of an HEB mixer. Two types of waveguide HEB devices have been used: a 0.8 THz HEB mixer and a 1.0 THz HEB mixer which is ~5 times smaller than the former. Two independent experimental approaches were used. In the ΔG/G method, the conversion gain of the HEB mixer is first measured as a function of the bias current for a number of bias voltages. At each bias setting, we carefully measure the change in the operating current when the input loads are switched. From the measured data, we can derive the expected difference in gain between the hot and cold loads. In the second method (injection method [1]), the linearity of the HEB mixer is independently measured by injecting a modulated signal for different input load temperatures. The results of both approaches confirm that there is gain compression in the operation of HEB mixers. Based on the results of our measurements, we discuss the impact of direct detection effects on the operation of HEB mixers.
|
Lobanov, Y. V., Tong, C. - Y. E., Hedden, A. S., Blundell, R., & Gol'tsman, G. N. (2010). Microwave-assisted슠measurement슠of the슠frequency슠response슠of슠terahertz슠HEB슠mixers슠with a슠fourier슠transform슠spectrometer. In 21st International Symposium on Space Terahertz Technology (pp. 420–423).
Abstract: We describe a novel method of operation of the HEB direct detector for use with a Fourier Transform Spectrometer. Instead of elevating the bath temperature, we have measured the RF response of waveguide HEB mixers by applying microwave radiation to select appropriate bias conditions. In our experiment, a microwave signal is injected into the HEB mixer via its IF port. By choosing an appropriate injection level, the device can be operated close to the desired operating point. Furthermore, we have shown that both thermal biasing and microwave injection can reproduce the same spectral response of the HEB mixer. However, with the use of microwave injection, there is no need to wait for the mixer to reach thermal equilibrium, so characterisation can be done in less time. Also, the liquid helium consumption for our wet cryostat is also reduced. We have demonstrated that the signalto-noise ratio of the FTS measurements can be improved with microwave injection.
|
Ожегов, Р. В., Окунев, О. В., & Гольцман, Г. Н. (2009). Флуктуационная чувствительность сверхпроводящего болометрического смесителя на эффекте разогрева электронного газа. Радиотехника, (3), 120–124.
Abstract: Interest in research in the terahertz range is driven by a great number of various applications, where terahertz instruments may play a leading role. To name just a few, such applications include study of the cosmic microwave background radiation and the distribution of the dark matter, medicine, navigation, fire alarm, security systems and environmental monitoring. The paper discusses the possibility of using a receiver based on the hot-electron effect in superconducting films as an imaging system. We present the results of the noise equivalent temperature difference (NETD) measurements performed with a hot-electron bolometer mixer made from a thin superconducting film. The receiver with a noise temperature of ~ 3800 K at a local oscillator frequency of 300 GHz a bandwidth of 500 MHz and an integration time of 1 s has offered an NETD of 0.5 K. We have also developed a technique that enabled us to reduce the contribution of the mixer gain fluctuations to the overall system instability. As of this writing, the above value of the NETD is the lowest value offered for this type of receiver, which indicates the possibility to use such receivers in real-time imaging systems. The technique offered in the paper for achieving the limiting value of the NETD offers an alternative to the phase-locking scheme.
Представены результаты измерения флуктуационной чувствительности (NETD – noise equivalent temperature difference) болометрического смесителя на эффекте разогрева электронного газа в тонких сверхпроводящих пленках. Получено предельное значение NETD, равное 0,5 К, при шумовой температуре приемника 3800 К, ширине полосы преобразования 500 МГц, постоянной времени 1 с и частоте гетеродина 300 ГГц. Разработана методика достижения предельной флуктуационной чувствительности, позволяющая избежать влияния нестабильности коэффициента преобразования смесителя.
|
Ожегов, Р. В. (2011). Флуктуационная чувствительность и стабильность приемников с СИС и HEB смесителями для терагерцового тепловидения. Радиофизика, , 135.
|
Shurakov, A., Tong, E., Blundell, R., & Gol'tsman, G. (2012). Microwave stabilization of HEB mixer by a microchip controller. In IEEE MTT-S international microwave symposium digest (pp. 1–3).
Abstract: The stability of a Hot Electron Bolometer (HEB) mixer can be improved by the use of microwave injection. In this article we report a refinement of this approach. We introduce a microchip controller to facilitate the implementation of the stabilization scheme, and demonstrate that the feedback loop effectively suppresses drifts in the HEB bias current, leading to an improvement in the receiver stability. The measured Allan time of the mixer's IF output power is increased to > 10 s.
|
Kawamura, J., Blundell, R., Tong, C. - Y. E., Golts'man, G., Gershenzon, E., & Voronov B. (1996). Superconductive NbN hot-electron bolometric mixer performance at 250 GHz. In Proc. 7th Int. Symp. Space Terahertz Technol. (pp. 331–336).
Abstract: Thin film NbN (<40 A) strips are used as waveguide mixer elements. The electron cooling mechanism for the geometry is the electron-phonon interaction. We report a receiver noise temperature of 750 K at 244 GHz, with / IF = 1.5 GHz, Af= 500 MHz, and Tphysical = 4 K. The instantaneous bandwidth for this mixer is 1.6 GHz. The local oscillator (LO) power is 0.5 1.tW with 3 dB-uncertainty. The mixer is linear to 1 dB up to an input power level 6 dB below the LO power. We report the first detection of a molecular line emission using this class of mixer, and that the receiver noise temperature determined from Y-factor measurements reflects the true heterodyne sensitivity.
|
Bell, M., Sergeev, A., Mitin, V., Bird, J., Verevkin, A., & Gol'tsman, G. (2007). One-dimensional resistive states in quasi-two-dimensional superconductors. arXiv:0709.0709v1 [cond-mat.supr-con], , 1–11.
Abstract: We investigate competition between one- and two-dimensional topological excitations – phase slips and vortices – in formation of resistive states in quasi-two-dimensional superconductors in a wide temperature range below the mean-field transition temperature T(C0). The widths w = 100 nm of our ultrathin NbN samples is substantially larger than the Ginzburg-Landau coherence length ξ = 4nm and the fluctuation resistivity above T(C0) has a two-dimensional character. However, our data shows that the resistivity below T(C0) is produced by one-dimensional excitations, – thermally activated phase slip strips (PSSs) overlapping the sample cross-section. We also determine the scaling phase diagram, which shows that even in wider samples the PSS contribution dominates over vortices in a substantial region of current/temperature variations. Measuring the resistivity within seven orders of magnitude, we find that the quantum phase slips can only be essential below this level.
|
Корнеева, Ю. П., Флоря, И. Н., Корнеев, А. А., & Гольцман, Г. Н. (2010). Cверхпроводящий однофотонный детектор для дальнего ИК диапазона длин волн. In Науч. сессия НИЯУ МИФИ (pp. 46–47).
Abstract: Мы представляем быстродействующий сверхпроводниковый однофотонный детектор (SSPD) для дальнего инфракрасного диапазона на основе ультратонкой монокристаллической пленки NbN толщиной 3 нм, состоящий из параллельных полосок. QE на длине волны 1,5.μм и 1,3 μм для предложенного SSPD практически одинаковы. SSPD показывает отклик длительностью 200 пс, что открывает путь к детекторам, обладающим скоростью счета свыше 1 ГГц.
|
Verevkin, A., Zhang, J., Slysz, W., Sobolewski, R., Lipatov, A., Okunev, O., et al. (2002). Spectral sensitivity and temporal resolution of NbN superconducting single-photon detectors. In Proc. 13th Int. Symp. Space Terahertz Technol. (pp. 105–111).
Abstract: We report our studies on spectral sensitivity and time resolution of superconducting NbN thin film single-photon detectors (SPDs). Our SPDs exhibit an everimentally measured detection efficiencies (DE) from — 0.2% at 2=1550 nm up to —3% at lambda=405 nm wavelength for 10-nm film thickness devices and up to 3.5% at lambda=1550 nm for 3.5-nm film thickness devices. Spectral dependences of detection efficiency (DE) at 2=0.4 —3.0 pm range are presented. With variable optical delay setup, it is shown that NbN SPD potentially can resolve optical pulses with the repetition rate up to 10 GHz at least. The observed full width at the half maximum (FWHM) of the signal pulse is about 150-180 ps, limited by read-out electronics. The jitter of NbN SPD is measured to be —35 ps at optimum biasing.
|
Beck, M., Klammer, M., Rousseau, I., Gol’tsman, G. N., Diamant, I., Dagan, Y., et al. (2015). Probing superconducting gap dynamics with THz pulses. In CLEO (SM3H.3 (1 to 2)). Optical Society of America.
Abstract: We studied superconducting gap dynamics in a BCS superconductor NbN and electron doped cuprate superconductor PCCO following excitation with near-infrared (NIR) and narrow band THz pulses. Systematic studies on PCCO imply very selective electron-phonon coupling.
|