Kooi JW, Baselmans JJA, Hajenius M, Gao JR, Klapwijk TM, Dieleman P, et al. IF impedance and mixer gain of NbN hot electron bolometers. J. Appl. Phys.. 2007;101(4):044511.
|
|
|
Kooi JW, Baselmans JJA, Baryshev A, Schieder R, Hajenius M, Gao JR, et al. Stability of heterodyne terahertz receivers. J Appl Phys. 2006;100(6):064904 (1 to 9).
Abstract: In this paper we discuss the stability of heterodyne terahertz receivers based on small volume NbN phonon cooled hot electron bolometers (HEBs). The stability of these receivers can be broken down in two parts: the intrinsic stability of the HEB mixer and the stability of the local oscillator (LO) signal injection scheme. Measurements show that the HEB mixer stability is limited by gain fluctuations with a 1∕f spectral distribution. In a 60MHz noise bandwidth this results in an Allan variance stability time of ∼0.3s. Measurement of the spectroscopic Allan variance between two intermediate frequency (IF) channels results in a much longer Allan variance stability time, i.e., 3s between a 2.5 and a 4.7GHz channel, and even longer for more closely spaced channels. This implies that the HEB mixer 1∕f noise is strongly correlated across the IF band and that the correlation gets stronger the closer the IF channels are spaced. In the second part of the paper we discuss atmospheric and mechanical system stability requirements on the LO-mixer cavity path length. We calculate the mixer output noise fluctuations as a result of small perturbations of the LO-mixer standing wave, and find very stringent mechanical and atmospheric tolerance requirements for receivers operating at terahertz frequencies.
|
|
|
Hajenius M, Baselmans JJA, Baryshev A, Gao JR, Klapwijk TM, Kooi JW, et al. Full characterization and analysis of a terahertz heterodyne receiver based on a NbN hot electron bolometer. J. Appl. Phys.. 2006;100(7):074507.
|
|
|
Datesman AM, Schultz JC, Lichtenberger AW, Golish D, Walker CK, Kooi J. Fabrication and characterization of niobium diffusion-cooled hot-electron bolometers on silicon nitride membranes. IEEE Trans. Appl. Supercond.. 2005;15(2):928–31.
|
|
|
Crockett NR, Bergin EA, Wang S, Lis DC, Bell TA, Blake GA, et al. Herschel observations of EXtra-Ordinary Sources (HEXOS): The Terahertz spectrum of Orion KL seen at high spectral resolution. Annual Rev. Astron. Astrophys.. 2010;521:L21 (1 to 5).
Abstract: We present the first high spectral resolution observations of Orion KL in the frequency ranges 1573.4–1702.8 GHz (band 6b) and 1788.4–1906.8 GHz (band 7b) obtained using the HIFI instrument on board the Herschel Space Observatory. We characterize the main emission lines found in the spectrum, which primarily arise from a range of components associated with Orion KL including the hot core, but also see widespread emission from components associated with molecular outflows traced by H2O, SO2, and OH. We find that the density of observed emission lines is significantly diminished in these bands compared to lower frequency Herschel/HIFI bands.
|
|
|
Baselmans J, Kooi J, Baryshev A, Yang ZQ, Hajenius M, Gao JR, et al. Full characterization of small volume NbN HEB mixers for space applications. In: Proc. 16th Int. Symp. Space Terahertz Technol. Göteborg, Sweden; 2005. p. 457–62.
Abstract: NbN phonon cooled HEB’s are one of the most promising bolometer mixer technologies for (near) future (space) applications. Their performance is usually quantified by mea- suring the receiver noise temperature at a given IF frequency, usually around 1 – 2 GHz. However, for any real applications it is vital that one fully knows all the relevant properties of the mixer, including LO power, stability, direct detection, gain bandwidth and noise bandwidth, not only the noise temperature at low IF frequencies. To this aim we have measured all these parameters at the optimal operating point of one single, small volume quasioptical NbN HEB mixer. We find a minimum noise temperature of 900 K at 1.46 THz. We observe a direct detection effect indicated by a change in bias current when changing from a 300 K hot load to a 77 K cold load. Due to this effect we overestimate the noise temperature by about 22% using a 300 K hot load and a 77 K cold load. The LO power needed to reach the optimal operating point is 80 nW at the receiver lens front, 59 nW inside the NbN bridge. However, using the isothermal technique we find a power absorbed in the NbN bridge of 25 nW, a difference of about a factor 2. We obtain a gain bandwidth of 2.3 GHz and a noise bandwidth of 4 GHz. The system Allan time is about 1 sec. in a 50 MHz spectral bandwidth and a deviation from white noise integration (governed by the radiometer equation) occurs at 0.2 sec., which implies a maximum integration time of a few seconds in a 1 MHz bandwidth spectrometer.
|
|
|
Baselmans JJA, Hajenius M, Gao JR, Baryshev A, Kooi J, Klapwijk TM, et al. NbN hot electron bolometer mixers: sensitivity, LO power, direct detection and stability. IEEE Trans Appl Supercond. 2005;15(2):484–9.
Abstract: We demonstrate that the performance of NbN lattice cooled hot electron bolometer mixers depends strongly on the interface quality between the bolometer and the contact structure. Both the receiver noise temperature and the gain bandwidth can be improved by a factor of 2 by cleaning the interface and adding an additional superconducting interlayer to the contact pad. Using this we obtain a double sideband receiver noise temperature of 950 K at 2.5 THz and 4.3 K, using a 0.4/spl times/4 /spl mu/m HEB mixer with a spiral antenna. At the same bias point, we obtain an IF gain bandwidth of 6 GHz. To comply with current demands on THz mixers for use in space based receivers we reduce the device size to 0.15/spl times/1 /spl mu/m and use a twin slot antenna. We report measurements of the noise temperature, LO power requirement, stability and the direct detection effect, using a mixer with a 1.6 THz twin slot antenna and a 1.462 THz solid state LO source with calibrated output power.
|
|
|
Baselmans JJA, Hajenius M, Gao JR, Baryshev A, Kooi J, Klapwijk TM, et al. Hot electron bolometer mixers with improved interfaces: sensitivity, LO power and stability. In: Proc. 15th Int. Symp. Space Terahertz Technol.; 2004. p. 17–24.
Abstract: We study twin slot antenna coupled NbN hot electron bolometer mixers with an improved contact structure and a small volume, ranging from 1 µm × 0.1 µm to 2 × 0.3 µm. We obtain a DSB receiver noise temperature of 900 K at 1.6 THz and 940 K at 1.9 THz. To explore the practical usability of such small HEB mixers we evaluate the LO power requirement, the sensitivity and the stability. We find that the LO power requirement of the smallest mixers is reduced to about 240 nW at the Si lens of the mixer. This value is larger than expected from the isothermal technique and the known losses in the lens by a factor of 3-3.5. The stability of these receivers is characterized using a measurement of the Allan Variance. We find an Allan time of 0.5 sec. in an 80 MHz bandwidth. A small increase in stability can be reached by using a higher bias at the expense of a significant amount of sensitivity. The stability is sufficient for spectroscopic applications in a 1 MHz bandwidth at a 1 Hz chopping frequency.
|
|