Ozhegov R, Maslennikov S, Morozov D, Okunev O, Smirnov K, Gol'tsman G. Imaging system for submillimeter wave range. In: Proc. Tenth All-Russian sceintific conference of student-physicists and young sceintists (VNKSF-10). Moscow; 2004.
|
|
|
Maslennikov S, Vachtomin Y, Antipov S, Smirnov K, Kaurova N, Grishina E, et al. NbN HEB mixers for frequencies of 2.5 and 3.8 THz. In: Proc. Tenth All-Russian sceintific conference of student-physicists and young sceintists (VNKSF-10). Moscow; 2004.
|
|
|
Antipov SV, Vachtomin YB, Maslennikov SN, Smirnov KV, Kaurova NS, Grishina EV, et al. Noise performance of quasioptical ultrathin NbN hot electron bolometer mixer at 2.5 and 3.8 THz. In: Proc. 5-th MSMW. Vol 2. Kharkov, Ukraine; 2004. p. 592–4.
Abstract: To put space-based and airborne heterodyne instruments into operation at frequencies above 1 THz the superconducting NbN hot-electron bolometer (HEB) will be incorporated into heterodyne receiver as a mixer. At frequencies above 1.3 THz the sensitivity of the NbN HEB mixers outperform the one of the Schottky diodes and SIS-mixers, and the receiver noise temperature of the NbN HEB mixers increase with frequency. In this paper we present the results of the noise temperature measurements within one batch of NbN HEB mixers based on 3.5 mn thick superconducting NbN film grown on Si substrate with MgO buffer layer at the LO frequencies 2.5 THz and 3.8 THz.
|
|
|
Vachtomin YB, Antipov SV, Kaurova NS, Maslennikov SN, Smirnov KV, Polyakov SL, et al. Noise temperature, gain bandwidth and local oscillator power of NbN phonon-cooled HEB mixer at terahertz frequenciess. In: Proc. 29th IRMMW / 12th THz. Karlsruhe, Germany; 2004. p. 329–30.
Abstract: We present the performances of HEB mixers based on 3.5 nm thick NbN film integrated with log-periodic spiral antenna. The double side-band receiver noise temperature values are 1300 K and 3100 K at 2.5 THz and at 3.8 THz, respectively. The gain bandwidth of the mixer is 4.2 GHz and the noise bandwidth is 5 GHz. The local oscillator power is 1-3 /spl mu/W for mixers with different active area.
|
|
|
Kaurova NS, Finkel MI, Maslennikov SN, Vahtomin YB, Antipov SV, Smirnov KV, et al. Submillimeter mixer based on YBa2Cu3O7-x thin film. In: Proc. 1-st conf. Fundamental problems of high temperature superconductivity. Moscow-Zvenigorod; 2004. 291.
|
|
|
Semenov A, Richter H, Smirnov K, Voronov B, Gol'tsman G, Hübers H-W. The development of terahertz superconducting hot-electron bolometric mixers. Supercond Sci Technol. 2004;17(5):436–9.
Abstract: We present recent advances in the development of NbN hot-electron bolometric (HEB) mixers for flying terahertz heterodyne receivers. Three important issues have been addressed: the quality of the source NbN films, the effect of the bolometer size on the spectral properties of different planar feed antennas, and the local oscillator (LO) power required for optimal operation of the mixer. Studies of the NbN films with an atomic force microscope indicated a surface structure that may affect the performance of the smallest mixers. Measured spectral gain and noise temperature suggest that at frequencies above 2.5 THz the spiral feed provides better overall performance than the double-slot feed. Direct measurements of the optimal LO power support earlier estimates made in the framework of the uniform mixer model.
|
|
|
Finkel MI, Maslennikov SN, Vachtomin YB, Svechnikov SI, Smirnov KV, Seleznev VA, et al. Hot electron bolometer mixer for 20 – 40 THz frequency range. In: Proc. 16th Int. Symp. Space Terahertz Technol. Göteborg, Sweden; 2005. p. 393–7.
Abstract: The developed HEB mixer was based on a 5 nm thick NbN film deposited on a GaAs substrate. The active area of the film was patterned as a 30×20 μm 2 strip and coupled with a 50 Ohm coplanar line deposited in situ. An extended hemispherical germanium lens was used to focus the LO radiation on the mixer. The responsivity of the mixer was measured in a direct detection mode in the 25÷64 THz frequency range. The noise performance of the mixer and the directivity of the receiver were investigated in a heterodyne mode. A 10.6 μm wavelength CW CO 2 laser was utilized as a local oscillator.
|
|
|
Gol'tsman GN, Vachtomin YB, Antipov SV, Finkel MI, Maslennikov SN, Smirnov KV, et al. NbN phonon-cooled hot-electron bolometer mixer for terahertz heterodyne receivers. In: Proc. SPIE. Vol 5727.; 2005. p. 95–106.
Abstract: We present the results of our studies of NbN phonon-cooled HEB mixers at terahertz frequencies. The mixers were fabricated from NbN film deposited on a high-resistivity Si substrate with an MgO buffer layer. The mixer element was integrated with a log-periodic spiral antenna. The noise temperature measurements were performed at 2.5 THz and at 3.8 THz local oscillator frequencies for the 3 x 0.2 μm2 active area devices. The best uncorrected receiver noise temperatures found for these frequencies are 1300 K and 3100 K, respectively. A water vapour discharge laser was used as the LO source. The largest gain bandwidth of 5.2 GHz was achieved for a mixer based on 2 nm thick NbN film deposited on MgO layer over Si substrate. The gain bandwidth of the mixer based on 3.5 nm NbN film deposited on Si with MgO is 4.2 GHz and the noise bandwidth for the same device amounts to 5 GHz. We also present the results of our research into decrease of the direct detection contribution to the measured Y-factor and a possible error of noise temperature calculation. The use of a square nickel cell mesh as an IR-filter enabled us to avoid the effect of direct detection and measure apparent value of the noise temperature which was 16% less than that obtained using conventional black polyethylene IR-filter.
|
|
|
Korneev A, Minaeva O, Rubtsova I, Milostnaya I, Chulkova G, Voronov B, et al. Superconducting single-photon ultrathin NbN film detector. Quantum Electronics. 2005;35(8):698–700.
Abstract: Superconducting single-photon ultrathin NbN film detectors are studied. The development of manufacturing technology of detectors and the reduction of their operating temperature down to 2 K resulted in a considerable increase in their quantum efficiency, which reached in the visible region (at 0.56 μm) 30%—40%, i.e., achieved the limit determined by the absorption coefficient of the film. The quantum efficiency exponentially decreases with increasing wavelength, being equal to ~20% at 1.55 μm and ~0.02% at 5 μm. For the dark count rate of ~10-4s-1, the experimental equivalent noise power was 1.5×10-20 W Hz-1/2; it can be decreased in the future down to the record low value of 5×10-21 W Hz-1/2. The time resolution of the detector is 30 ps.
|
|
|
Maslennikov SN, Finkel MI, Antipov SV, Polyakov SL, Zhang W, Ozhegov R, et al. Spiral antenna coupled and directly coupled NbN HEB mixers in the frequency range from 1 to 70 THz. In: Proc. 17th Int. Symp. Space Terahertz Technol. Paris, France; 2006. p. 177–9.
Abstract: We investigate both antenna coupled and directly coupled HEB mixers at several LO frequencies within the range of 2.5 THz to 70 THz. H20 (2.5+10.7 THz), and CO2 (30 THz) gas discharge lasers are used as the local oscillators. The noise temperature of antenna coupled mixers is measured at LO frequencies of 2.5 THz, 3.8 THz, and 30 THz. The results for both antenna coupled and directly coupled mixer types are compared. The devices with in—plane dimensions of 5x5 ,um 2 are pumped by LO radiation at 10.7 THz. The directly coupled HEB demonstrates nearly flat dependence of responsivity on frequency in the range of 25+64 THz.
|
|