Maslennikov S, Antipov S, Shishkov A, Svechnikov S, Voronov B, Smirnov K, et al. NbN HEB mixer noise temperature measurements with hot/cold load mounted inside the helium cryostat at 300 GHz. In: Proc. Int. Student Seminar on Microwave Appl. of Novel Physical Phenomena supported by IEEE. St.-Petersburg: LETI; 2002.
|
|
|
Vahtomin YB, Finkel MI, Antipov SV, Voronov BM, Smirnov KV, Kaurova NS, et al. Gain bandwidth of phonon-cooled HEB mixer made of NbN thin film with MgO buffer layer on Si. In: Harvard university, editor. Proc. 13th Int. Symp. Space Terahertz Technol. Cambridge, MA, USA; 2002. p. 259–70.
Abstract: We present recently obtained values for gain bandwidth of NbN HEB mixers for different substrates and film thicknesses and for MgO buffer layer on Si at LO frequency of 0.85-1 THz. The maximal bandwidth, 5.2 GHz, was achieved for the device on MgO buffer layer on Si with a 2 nm thick NbN film. Functional devices based on NbN films of such thickness were fabricated for the first time due to an improvement of superconducting properties of NbN film deposited on MgO buffer layer on Si substrate.
|
|
|
Meledin D, Tong CY-E, Blundell R, Kaurova N, Smirnov K, Voronov B, et al. The sensitivity and IF bandwidth of waveguide NbN hot electron bolometer mixers on MgO buffer layers over crystalline quartz. In: Harvard university, editor. Proc. 13th Int. Symp. Space Terahertz Technol. Cambridge, MA, USA; 2002. p. 65–72.
Abstract: We have developed and characterized waveguide phonon-cooled NbN Hot Electron Bolometer (FMB) mixers fabricated from a 3-4 nm thick NbN film deposited on a 200nm thick MgO buffer layer over crystalline quartz. Double side band receiver noise temperatures of 900-1050 K at 1.035 THz, and 1300-1400 K at 1.26 THz have been measured at an intermediate frequency of 1.5 GHz. The intermediate frequency bandwidth, measured at 0.8 THz LO frequency, is 3.2 GHz at the optimal bias point for low noise receiver operation.
|
|
|
Huebers H-W, Semenov A, Richter H, Birk M, Krocka M, Mair U, et al. Superconducting hot electron bolometer as mixer for far-infrared heterodyne receivers. In: Phillips TG, Zmuidzinas J, editors. Proc. SPIE. Vol 4855. Tucson, USA: SPIE; 2003. p. 395–401. (Presented at the Society of Photo-Optical Instrumentation Engineers (SPIE) Conference; vol 4855).
Abstract: Heterodyne receivers for applications in astronomy need quantum limited sensitivity. In instruments which are currently under development for SOFIA or Herschel superconducting hot electron bolometers (HEB) will be used to achieve this goal at frequencies above 1.4 THz. We present results of the development of a phonon-cooled NbN HEB mixer for GREAT, the German Receiver for Astronomy at Terahertz Frequencies, which will be flown aboard SOFIA. The mixer is a small superconducting bridge incorporated in a planar feed antenna and a hyperhemispherical lens. Mixers with logarithmic-spiral and double-slot feed antennas have been investigated with respect to their noise temperature, conversion loss, linearity and beam pattern. At 2.5 THz a double sideband noise temperature of 2200 K was achieved. The conversion loss was 17 dB. The response of the mixer was linear up to 400 K load temperature. The performance was verified by measuring an emission line of methanol at 2.5 THz. The measured linewidth is in good agreement with the linewidth deduced from pressure broadening measurements at millimeter wavelength. The results demonstrate that the NbN HEB is very well suited as a mixer for far-infrared heterodyne receivers.
|
|
|
Vachtomin YB, Antipov SV, Maslennikov SN, Smirnov KV, Polyakov SL, Kaurova NS, et al. Noise temperature measurements of NbN phonon-cooled hot electron bolometer mixer at 2.5 and 3.8 THz. In: Proc. 15th Int. Symp. Space Terahertz Technol. Northampton, Massachusetts, USA; 2004. p. 236–41.
Abstract: We present the results of noise temperature measurements of NbN phonon-cooled HEB mixers based on a 3.5 nm NbN film deposited on a high-resistivity Si substrate with a 200 nm – thick MgO buffer layer. The mixer element was integrated with a log-periodic spiral antenna. The noise temperature measurements were performed at 2.5 THz and at 3.8 THz local oscillator frequencies for the 3 µm x 0.2 µm active area devices. The best uncorrected receiver noise temperatures found for these frequencies are 1300 K and 3100 K, respectively. A water vapour discharge laser was used as the LO source. We also present the results of direct detection contribution to the measured Y-factor and of a possible error of noise temperature calculation. This error was more than 8% for the mixer with in-plane dimensions of 2.4 x 0.16 µm 2 at the optimal noise temperature point. The use of a mesh filter enabled us to avoid the effect of direct detection and decrease optical losses by 0.5 dB. The paper is concluded by the investigation results of the mixer polarization response. It was shown that the polarization can differ from the circular one at 3.8 THz by more than 2 dB.
|
|
|
Ozhegov R, Morozov D, Maslennikov S, Okunev O, Smirnov K, Gol'tsman G. Submillimeter wave range imaging system for registering human body radiation and finding out the things covered under clothes. In: Proc. 3rd Int. exhibition and conf. Non-Destructive Testing Equipment and Devices. Moscow; 2004.
|
|
|
Ozhegov R, Maslennikov S, Morozov D, Okunev O, Smirnov K, Gol'tsman G. Imaging system for submillimeter wave range. In: Proc. Tenth All-Russian sceintific conference of student-physicists and young sceintists (VNKSF-10). Moscow; 2004.
|
|
|
Maslennikov S, Vachtomin Y, Antipov S, Smirnov K, Kaurova N, Grishina E, et al. NbN HEB mixers for frequencies of 2.5 and 3.8 THz. In: Proc. Tenth All-Russian sceintific conference of student-physicists and young sceintists (VNKSF-10). Moscow; 2004.
|
|
|
Antipov SV, Vachtomin YB, Maslennikov SN, Smirnov KV, Kaurova NS, Grishina EV, et al. Noise performance of quasioptical ultrathin NbN hot electron bolometer mixer at 2.5 and 3.8 THz. In: Proc. 5-th MSMW. Vol 2. Kharkov, Ukraine; 2004. p. 592–4.
Abstract: To put space-based and airborne heterodyne instruments into operation at frequencies above 1 THz the superconducting NbN hot-electron bolometer (HEB) will be incorporated into heterodyne receiver as a mixer. At frequencies above 1.3 THz the sensitivity of the NbN HEB mixers outperform the one of the Schottky diodes and SIS-mixers, and the receiver noise temperature of the NbN HEB mixers increase with frequency. In this paper we present the results of the noise temperature measurements within one batch of NbN HEB mixers based on 3.5 mn thick superconducting NbN film grown on Si substrate with MgO buffer layer at the LO frequencies 2.5 THz and 3.8 THz.
|
|
|
Vachtomin YB, Antipov SV, Kaurova NS, Maslennikov SN, Smirnov KV, Polyakov SL, et al. Noise temperature, gain bandwidth and local oscillator power of NbN phonon-cooled HEB mixer at terahertz frequenciess. In: Proc. 29th IRMMW / 12th THz. Karlsruhe, Germany; 2004. p. 329–30.
Abstract: We present the performances of HEB mixers based on 3.5 nm thick NbN film integrated with log-periodic spiral antenna. The double side-band receiver noise temperature values are 1300 K and 3100 K at 2.5 THz and at 3.8 THz, respectively. The gain bandwidth of the mixer is 4.2 GHz and the noise bandwidth is 5 GHz. The local oscillator power is 1-3 /spl mu/W for mixers with different active area.
|
|