Lobanov YV, Tong C-YE, Hedden AS, Blundell R, Voronov BM, Gol'tsman GN. Direct measurement of the gain and noise bandwidths of HEB mixers. IEEE Trans Appl Supercond. 2011;21(3):645–8.
Abstract: The intermediate frequency (IF) bandwidth of a hot electron bolometer (HEB) mixer is an important parameter of the mixer, in that it helps to determine its suitability for a given application. With the availability of wideband low noise amplifiers, it is simple to measure the performance of an HEB mixer over a wide range of IF at a fixed LO frequency using the standard Y-factor method. This in-situ method allows us to measure both the gain and noise bandwidths simultaneously. We have also measured mixer output impedance with a vector network analyser. Intrinsic time constant has been extracted from the impedance data and compared to the mixer's bandwidths determined from receiver Y-factor measurement.
|
Lobanov Y, Tong C, Blundell R, Gol'tsman G. A study of direct detection effect on the linearity of hot electron bolometer mixers. In: Proc. 20th Int. Symp. Space Terahertz Technol.; 2009. p. 282–7.
Abstract: We have performed a study of how direct detection affects the linearity and hence the calibration of an HEB mixer. Two types of waveguide HEB devices have been used: a 0.8 THz HEB mixer and a 1.0 THz HEB mixer which is ~5 times smaller than the former. Two independent experimental approaches were used. In the ΔG/G method, the conversion gain of the HEB mixer is first measured as a function of the bias current for a number of bias voltages. At each bias setting, we carefully measure the change in the operating current when the input loads are switched. From the measured data, we can derive the expected difference in gain between the hot and cold loads. In the second method (injection method [1]), the linearity of the HEB mixer is independently measured by injecting a modulated signal for different input load temperatures. The results of both approaches confirm that there is gain compression in the operation of HEB mixers. Based on the results of our measurements, we discuss the impact of direct detection effects on the operation of HEB mixers.
|
Kawamura J, Blundell R, Tong C-YE, Golts'man G, Gershenzon E, Voronov B. Superconductive NbN hot-electron bolometric mixer performance at 250 GHz. In: Proc. 7th Int. Symp. Space Terahertz Technol.; 1996. p. 331–6.
Abstract: Thin film NbN (<40 A) strips are used as waveguide mixer elements. The electron cooling mechanism for the geometry is the electron-phonon interaction. We report a receiver noise temperature of 750 K at 244 GHz, with / IF = 1.5 GHz, Af= 500 MHz, and Tphysical = 4 K. The instantaneous bandwidth for this mixer is 1.6 GHz. The local oscillator (LO) power is 0.5 1.tW with 3 dB-uncertainty. The mixer is linear to 1 dB up to an input power level 6 dB below the LO power. We report the first detection of a molecular line emission using this class of mixer, and that the receiver noise temperature determined from Y-factor measurements reflects the true heterodyne sensitivity.
|
Trifonov A, Tong C-YE, Blundell R, Ryabchun S, Gol'tsman G. Probing the stability of HEB mixers with microwave injection. IEEE Trans Appl Supercond. 2015;25(3):2300404 (1 to 4).
Abstract: Using a microwave probe as a tool, we have performed experiments aimed at understanding the origin of the output-power fluctuations in hot-electron-bolometer (HEB) mixers. We use a probe frequency of 1.5 GHz. The microwave probe picks up impedance changes of the HEB, which are examined upon demodulation of the reflected wave outside the cryostat. This study shows that the HEB mixer operates in two different regimes under a terahertz pump. At a low pumping level, strong pulse modulation is observed, as the device switches between the superconducting state and the normal state at a rate of a few megahertz. When pumped much harder, to approximate the low-noise mixer operating point, residual modulation can still be observed, showing that the HEB mixer is intrinsically unstable even in the resistive state. Based on these observations, we introduced a low-frequency termination to the HEB mixer. By terminating the device in a 50-Ω resistor in the megahertz frequency range, we have been able to improve the output-power Allan time of our HEB receiver by a factor of four to about 10 s for a detection bandwidth of 15 MHz, with a corresponding gain fluctuation of about 0.035%.
|
Trifonov A, Tong C-YE, Grimes P, Lobanov Y, Kaurova N, Blundell R, et al. Development of A Silicon Membrane-based Multi-pixel Hot Electron Bolometer Receiver. In: IEEE Trans. Appl. Supercond. Vol 27.; 2017. 6.
Abstract: We report on the development of a multi-pixel
Hot Electron Bolometer (HEB) receiver fabricated using
silicon membrane technology. The receiver comprises a
2 × 2 array of four HEB mixers, fabricated on a single
chip. The HEB mixer chip is based on a superconducting
NbN thin film deposited on top of the silicon-on-insulator
(SOI) substrate. The thicknesses of the device layer and
handling layer of the SOI substrate are 20 μm and 300 μm
respectively. The thickness of the device layer is chosen
such that it corresponds to a quarter-wave in silicon at
1.35 THz. The HEB mixer is integrated with a bow-tie
antenna structure, in turn designed for coupling to a
circular waveguide,
|