Author Shurakov, Alexander; Tong, Edward; Blundell, Raymond; Gol'tsman, Gregory
Title Microwave stabilization of HEB mixer by a microchip controller Type Conference Article
Year 2012 Publication IEEE MTT-S international microwave symposium digest Abbreviated Journal
Volume (up) Issue Pages 1-3
Keywords HEB mixer stability, microwave injection, Allan variance, Allan time
Abstract The stability of a Hot Electron Bolometer (HEB) mixer can be improved by the use of microwave injection. In this article we report a refinement of this approach. We introduce a microchip controller to facilitate the implementation of the stabilization scheme, and demonstrate that the feedback loop effectively suppresses drifts in the HEB bias current, leading to an improvement in the receiver stability. The measured Allan time of the mixer's IF output power is increased to > 10 s.
Address Montreal, QC, Canada
Corporate Author Thesis
Publisher Place of Publication Editor
Language Summary Language Original Title
Series Editor Series Title Abbreviated Series Title
Series Volume Series Issue Edition
ISSN ISBN Medium
Area Expedition Conference
Notes Approved no
Call Number Serial 857
Permanent link to this record
 

 
Author Khosropanah, P.; Merkel, H.; Yngvesson, S.; Adam, A.; Cherednichenko, S.; Kollberg, E.
Title A distributed device model for phonon-cooled HEB mixers predicting IV characteristics, gain, noise and IF bandwidth Type Conference Article
Year 2000 Publication Proc. 11th Int. Symp. Space Terahertz Technol. Abbreviated Journal
Volume (up) Issue Pages 474-488
Keywords HEB mixer numerical model, diffusion cooling channel, diffusion channel, distributed HEB model, distributed model
Abstract A distributed model for phonon-cooled superconductor hot electron bolometer (HEB) mixers is given, which is based on solving the one-dimensional heat balance equation for the electron temperature profile along the superconductor strip. In this model it is assumed that the LO power is absorbed uniformly along the bridge but the DC power absorption depends on the local resistivity and is thus not uniform. The electron temperature dependence of the resistivity is assumed to be continuous and has a Fermi form. These assumptions are used in setting up the non-linear heat balance equation, which is solved numerically for the electron temperature profile along the bolometer strip. Based on this profile the resistance of the device and the IV curves are calculated. The IV curves are in excellent agreement with measurement results. Using a small signal model the conversion gain of the mixer is obtained. The expressions for Johnson noise and thermal fluctuation noise are derived. The calculated results are in close agreement with measurements, provided that one of the parameters used is adjusted.
Address
Corporate Author Thesis
Publisher Place of Publication University of Michigan, Ann Arbor, MI USA Editor
Language Summary Language Original Title
Series Editor Series Title Abbreviated Series Title
Series Volume Series Issue Edition
ISSN ISBN Medium
Area Expedition Conference
Notes Approved no
Call Number Serial 893
Permanent link to this record
 

 
Author Kooi, Jacob Willem
Title Advanced receivers for submillimeter and far infrared astronomy Type Book Whole
Year 2008 Publication University of Groningen Abbreviated Journal RUG
Volume (up) Issue Pages
Keywords HEB, SIS, TES, NEP, noise temperature, IF bandwidth, waveguide, impedance, conversion gain, FTS, integrated array, stability, Allan variance, multi-layer antireflection coating
Abstract
Address
Corporate Author Thesis Doctoral thesis
Publisher Place of Publication Editor
Language Summary Language Original Title
Series Editor Series Title Abbreviated Series Title
Series Volume Series Issue Edition
ISSN ISBN 978-90-367-3653-4 Medium
Area Expedition Conference
Notes Approved no
Call Number Serial 881
Permanent link to this record
 

 
Author Ynvesson, K. Sigfrid; Kollberg, Erik L.
Title Optimum receiver noise temperature for NbN HEB mixers according to standard model Type Conference Article
Year 1999 Publication Proc. 10th Int. Symp. Space Terahertz Technol. Abbreviated Journal
Volume (up) Issue Pages 566-582
Keywords HEB mixer model, standard model, electro-thermal feedback, self-heating parameter, heating efficiency
Abstract
Address
Corporate Author Thesis
Publisher Place of Publication Editor
Language Summary Language Original Title
Series Editor Series Title Abbreviated Series Title
Series Volume Series Issue Edition
ISSN ISBN Medium
Area Expedition Conference
Notes Approved no
Call Number Serial 895
Permanent link to this record
 

 
Author Khosropanah, Pourya
Title NbN and NbTiN hot electron bolometer THz mixers Type Book Whole
Year 2003 Publication Chalmers University of Technology Abbreviated Journal
Volume (up) Issue Pages
Keywords HEB mixer, hot electron bolometer mixer, NbN, NbTiN, superconducting detector, heterodyne receiver, THz mixer, submillimeter mixer, quasioptical receiver, double slot antenna, twin slot antenna, spiral antenna, receiver noise, FTS, Fourier Transform Spectrometer
Abstract The thesis reports the development of Hot Electron Bolometer (HEB) mixers for radio astronomy heterodyne receivers in THz frequency range. Part of this work is the fabrication of HEB devices, which are based on NbN or NbTiN superconducting thin films (â‰<a4>5 nm). They are integrated with wideband spiral or double-slot planar antennas. The mixer chips are incorporated into a quasi-optical receiver. The experimental part of this work focuses on the characterization of the receiver as a whole, and the HEB mixers as a part. Double side band receiver noise temperature and the IF bandwidth are reported for frequencies from 0.7 THz up to 2.6 THz. The spectrum of the direct response of HEB integrated with dierent antennas are measured using Fourier Transform Spectrometer (FTS). The effect of the bolometer size on total receiver performance and the LO power requirements is also discussed. A high-yield and reliable process for fabrication of NbN HEB mixers have been achieved. Over 100 devices with different bolometer geometry, film property and also different antennas have been fabricated and measured. The measured data enables us to discuss the impact of different parameters to the receiver overall performance.

This work has provided NbN HEB mixers to the following receivers:

TREND (Terahertz REceiver with NbN HEB Device) operating at 1.25-1.5 THz, installed in AST/RO Submillimeter Wave Telescope, Amundsen/Scott South Pole Station, in 2002-2003.

Band 6-low (1.410-1.700 THz) and 6-high (1.700-1.920 THz) of the HIFI (Heterodyne Instrument for Far Infra-red) in the Herschel Space Observatory, due to launch in 2007 by ESA (European Space Agency).

Besides, there has been continuous efforts to develop better models to explain the mixer performance more accurately. They are based on two temperature model for electrons and phonons and solving one-dimensional heat balance equations along the bolometer. The principles of these models are illustrated and the calculated results are compared with measured data.
Address
Corporate Author Thesis Ph.D. thesis
Publisher Chalmers University of Technology Place of Publication Göteborg Editor
Language Summary Language Original Title
Series Editor Series Title Abbreviated Series Title
Series Volume Series Issue Edition
ISSN ISBN Medium
Area Expedition Conference
Notes Approved no
Call Number Serial 910
Permanent link to this record