Maslennikov, S., Antipov, S., Shishkov, A., Svechnikov, S., Voronov, B., Smirnov, K., et al. (2002). NbN HEB mixer noise temperature measurements with hot/cold load mounted inside the helium cryostat at 300 GHz. In Proc. Int. Student Seminar on Microwave Appl. of Novel Physical Phenomena supported by IEEE. St.-Petersburg: LETI.
|
|
|
Maslennikov, S., Vachtomin, Y., Antipov, S., Smirnov, K., Kaurova, N., Grishina, E., et al. (2004). NbN HEB mixers for frequencies of 2.5 and 3.8 THz. In Proc. Tenth All-Russian sceintific conference of student-physicists and young sceintists (VNKSF-10). Moscow.
|
|
|
Smirnov, K. V., Vachtomin, Y. B., Antipov, S. V., Maslennikov, S. N., Kaurova, N. S., Drakinsky, V. N., et al. (2003). Noise and gain performance of spiral antenna coupled HEB mixers at 0.7 THz and 2.5 THz. In Proc. 14th Int. Symp. Space Terahertz Technol. (pp. 405–412).
Abstract: Noise and gain performance of hot electron bolometer (HEB) mixers based on ultrathin superconducting NbN films integrated with a spiral antenna was studied. The noise temperature measurements for two samples with different active area of 3 p.m x 0.24 .tni and 1.3 1..tm x 0.12 1.tm were performed at frequencies 0.7 THz and 2.5 THz. The best receiver noise temperatures 370 K and 1600 K, respectively, have been found at these frequencies. The influence of contact resistance between the superconductor and the antenna terminals on the noise temperature of HEB is discussed. The noise and gain bandwidth of 5GHz and 4.2 GHz, respectively, are demonstrated for similar HEB mixer at 0.75 THz.
|
|
|
Zhang, W., Li, N., Jiang, L., Miao, W., Lin, Z. - H., Yao, Q. - J., et al. (2007). Noise behaviour of a THz superconducting hot-electron bolometer mixer. Chinese Phys. Lett., 24(6), 1778–1781.
Abstract: A quasi-optical superconducting NbN hot-electron bolometer (HEB) mixer is measured in the frequency range of 0.5–2.5 THz for understanding of the frequency dependence of noise temperature of THz coherent detectors. It has been found that noise temperature increasing with frequency is mainly due to the coupling loss between the quasi-optical planar antenna and the superconducting HEB bridge when taking account of non-uniform distribution of high-frequency current. With the coupling loss corrected, the superconducting HEB mixer demonstrates a noise temperature nearly independent of frequency.
|
|
|
Antipov, S. V., Vachtomin, Y. B., Maslennikov, S. N., Smirnov, K. V., Kaurova, N. S., Grishina, E. V., et al. (2004). Noise performance of quasioptical ultrathin NbN hot electron bolometer mixer at 2.5 and 3.8 THz. In Proc. 5-th MSMW (Vol. 2, pp. 592–594). Kharkov, Ukraine.
Abstract: To put space-based and airborne heterodyne instruments into operation at frequencies above 1 THz the superconducting NbN hot-electron bolometer (HEB) will be incorporated into heterodyne receiver as a mixer. At frequencies above 1.3 THz the sensitivity of the NbN HEB mixers outperform the one of the Schottky diodes and SIS-mixers, and the receiver noise temperature of the NbN HEB mixers increase with frequency. In this paper we present the results of the noise temperature measurements within one batch of NbN HEB mixers based on 3.5 mn thick superconducting NbN film grown on Si substrate with MgO buffer layer at the LO frequencies 2.5 THz and 3.8 THz.
|
|