Ren, Y., Zhang, D. X., Zhou, K. M., Miao, W., Zhang, W., Shi, S. C., et al. (2019). 10.6 μm heterodyne receiver based on a superconducting hot-electron bolometer mixer and a quantum cascade laser. AIP Advances, 9(7), 075307.
Abstract: We report on the development of a heterodyne receiver at mid-infrared wavelength for high-resolution spectroscopy applications. The receiver employs a superconducting NbN hot electron bolometer as a mixer and a room temperature distributed feedback quantum cascade laser operating at 10.6 μm (28.2 THz) as a local oscillator. The stabilization of the heterodyne receiver has been achieved using a feedback loop controlling the output power of the laser. Improved Allan variance times as well as a double sideband receiver noise temperature of 5000 K and a noise bandwidth of 2.8 GHz of the receiver system are demonstrated.
The work is supported in part by the National Key R&D Program of China under Grant 2018YFA0404701, by the CAS program under Grant QYZDJ-SSW-SLH043 and GJJSTD20180003, by the National Natural Science Foundation of China (NSFC) under Grant 11773083, by the “Hundred Talents Program” of the “Pioneer Initiative”, and in part by the CAS Key Lab for Radio Astronomy.
|
Jiang, L., Zhang, W., Yao, Q. J., Lin, Z. H., Li, J., Shi, S. C., et al. (2005). Characterization of a quasi-optical NbN superconducting hot-electron bolometer mixer. In Proc. PIERS (Vol. 1, pp. 587–590).
Abstract: In this paper, we report the performance of a quasi-optical NbN superconducting HEB (hot electron bolome-ter) mixer measured at 500 GHz. The quasi-optical NbN superconducting HEB mixer is cryogenically cooled bya 4-K close-cycled refrigerator. Its receiver noise temperature and conversion gain are thoroughly investigatedfor different LO pumping levels and dc biases. The lowest receiver noise temperature is found to be approxi-mately 1200 K, and reduced to about 445 K after correcting theloss of the measurement system. The stabilityof the mixer’s IF output power is also demonstrated.
|
Jiang, L., Li, J., Zhang, W., Yao, Q. J., Lin, Z. L., Shi, S. C., et al. (2005). Characterization of NbN HEB mixers cooled by a close-cycled 4 Kelvin refrigerator. IEEE Trans. Appl. Supercond., 15(2), 511–513.
Abstract: It is quite beneficial to operate superconducting hot-electron-bolometer (HEB) mixers with a close-cycled 4 Kelvin refrigerator for real applications such as astronomy and atmospheric research. In this paper, a phononcooled NbN HEB mixer (quasioptical type) is thoroughly characterized under such a cooling circumstance. The effects of mechanical vibration, electrical interference, and temperature fluctuation of a two-stage Gifford-McMahon 4 Kelvin refrigerator upon the characteristics of the phononcooled NbN HEB mixer are investigated in particular. Detailed measurement results are presented.
|
Jiang, L., Antipov, S. V., Voronov, B. M., Gol'tsman, G. N., Zhang, W., Li, N., et al. (2007). Characterization of the performance of a quasi-optical NbN superconducting HEB mixer. IEEE Trans. Appl. Supercond., 17(2), 395–398.
Abstract: In this paper we focus mainly on the investigation of the performance of a quasi-optical (planar log-spiral antenna) phonon-cooled NbN superconducting hot electron bolometer (HEB) mixer, which is cryogenically cooled by a close-cycled 4-K cryocooler, at 500 and 850 GHz frequency bands. The mixer's noise performance, stability of IF output power, and local oscillator (LO) power requirement are characterized for three NbN superconducting HEB devices of different sizes. The transmission characteristics of Mylar and Zitex films with incidence waves of an elliptical polarization are also examined by measuring the mixer's noise temperature. The lowest receiver noise temperatures (with no corrections) of 750 and 1100 K are measured at 500 and 850 GHz, respectively. Experimental results also demonstrate that the bigger the HEB device is, the higher the stability of IF output power becomes.
|
Shi, S. C., Chin, C. C., Wang, M. J., Shan, W. L., Zhang, W., & Noguchi, T. (2001). Development of a 600–720 GHz SIS Mixer for the SMART. In C. Iit.u.t.e of T. Jet Propulsion Laboratory (Ed.), Proc. 12th Int. Symp. Space Terahertz Technol. (215). San Diego, CA, USA.
|