Jiang, L., Li, J., Zhang, W., Yao, Q. J., Lin, Z. L., Shi, S. C., et al. (2005). Characterization of NbN HEB mixers cooled by a close-cycled 4 Kelvin refrigerator. IEEE Trans. Appl. Supercond., 15(2), 511–513.
Abstract: It is quite beneficial to operate superconducting hot-electron-bolometer (HEB) mixers with a close-cycled 4 Kelvin refrigerator for real applications such as astronomy and atmospheric research. In this paper, a phononcooled NbN HEB mixer (quasioptical type) is thoroughly characterized under such a cooling circumstance. The effects of mechanical vibration, electrical interference, and temperature fluctuation of a two-stage Gifford-McMahon 4 Kelvin refrigerator upon the characteristics of the phononcooled NbN HEB mixer are investigated in particular. Detailed measurement results are presented.
|
|
|
Chen, J., Kang, L., Jin, B. B., Xu, W. W., Wu, P. H., Zhang, W., et al. (2008). Properties of terahertz superconducting hot electron bolometer mixers. Int. J. Terahertz Sci. Technol., 1(1), 37–41.
Abstract: A quasi-optical superconducting niobium nitride (NbN) hot electron bolometer (HEB) mixer has been fabricated and measured in the terahertz (THz) frequency range of 0.5~2.52 THz. A receiver noise temperature of 2000 K at 2.52 THz has been obtained for the mixer without corrections. Also, the effect of a Parylene C anti-reflection (AR) coating on the silicon (Si) lens has been studied.
|
|
|
Jiang, L., Zhang, W., Yao, Q. J., Lin, Z. H., Li, J., Shi, S. C., et al. (2005). Characterization of a quasi-optical NbN superconducting hot-electron bolometer mixer. In Proc. PIERS (Vol. 1, pp. 587–590).
Abstract: In this paper, we report the performance of a quasi-optical NbN superconducting HEB (hot electron bolome-ter) mixer measured at 500 GHz. The quasi-optical NbN superconducting HEB mixer is cryogenically cooled bya 4-K close-cycled refrigerator. Its receiver noise temperature and conversion gain are thoroughly investigatedfor different LO pumping levels and dc biases. The lowest receiver noise temperature is found to be approxi-mately 1200 K, and reduced to about 445 K after correcting theloss of the measurement system. The stabilityof the mixer’s IF output power is also demonstrated.
|
|
|
Zhang, W., Miao, W., Li, S. L., Zhou, K. M., Shi, S. C., Gao, J. R., et al. (2013). Measurement of the spectral response of spiral-antenna coupled superconducting hot electron bolometers. IEEE Trans. Appl. Supercond., 23(3), 2300804.
Abstract: Measured spectral response of spiral-antenna coupled superconducting hot electron bolometers (HEBs) often drops dramatically at frequencies that are still within the frequency range of interest (e.g., ~ 5 THz). This is inconsistent with the implied low receiver noise temperatures from the same measurements. To understand this discrepancy, we exhaustively test and calibrate the thermal sources used in Fourier transform spectrometer measurements. We first investigate the absolute emission spectrum of high-pressure Hg arc lamp, then measure the spectral response of two spiral-antenna coupled NbN HEBs with a Martin-Puplett interferometer as spectrometer and 77 K blackbody as broadband signal source. The measured absolute emission spectrum of Hg arc lamp is proportional to frequency, corresponding to an equivalent blackbody temperature of 4000 K at 1 THz, 1500 K at 3 THz, and 800 K at 5 THz, respectively. Measured spectral response of spiral-antenna coupled NbN HEBs, corrected for air absorption, is nearly flat in the frequency range of 0.5-4 THz, consistent with simulated coupling efficiency between HEB and spiral-antenna. These results explain the discrepancy, and prove that spiral-antenna coupled superconducting NbN HEBs work well in a wide frequency range. In addition, this calibration method and these results are broadly applicable to other quasi-optical THz receivers.
|
|
|
Miao, W., Zhang, W., Zhong, J. Q., Shi, S. C., Delorme, Y., Lefevre, R., et al. (2014). Non-uniform absorption of terahertz radiation on superconducting hot electron bolometer microbridges. <ef><bf><bc>Appl. Phys. Lett., 104, 052605(1–4).
Abstract: We interpret the experimental observation of a frequency-dependence of superconducting hot electron bolometer (HEB) mixers by taking into account the non-uniform absorption of the terahertz radiation on the superconducting HEB microbridge. The radiation absorption is assumed to be proportional to the local surface resistance of the HEB microbridge, which is computed using the Mattis-Bardeen theory. With this assumption the dc and mixing characteristics of a superconducting niobium-nitride (NbN) HEB device have been modeled at frequencies below and above the equilibrium gap frequency of the NbN film.
|
|