Finkel, M. I., Maslennikov, S. N., Vachtomin, Y. B., Svechnikov, S. I., Smirnov, K. V., Seleznev, V. A., et al. (2005). Hot electron bolometer mixer for 20 – 40 THz frequency range. In Proc. 16th Int. Symp. Space Terahertz Technol. (pp. 393–397). Göteborg, Sweden.
Abstract: The developed HEB mixer was based on a 5 nm thick NbN film deposited on a GaAs substrate. The active area of the film was patterned as a 30×20 μm 2 strip and coupled with a 50 Ohm coplanar line deposited in situ. An extended hemispherical germanium lens was used to focus the LO radiation on the mixer. The responsivity of the mixer was measured in a direct detection mode in the 25÷64 THz frequency range. The noise performance of the mixer and the directivity of the receiver were investigated in a heterodyne mode. A 10.6 μm wavelength CW CO 2 laser was utilized as a local oscillator.
|
|
|
Loudkov, D., Tong, C. - Y. E., Marrone, D. P., Ryabchun, S., Paine, S. N., & Blundell, R. (2005). Transmission measurements of infrared filters for low-noise terahertz receiver applications. In Proc. 16th Int. Symp. Space Terahertz Technol. (pp. 354–357).
Abstract: Infrared (IR) filters are very important to the efficient operation of cryogenic receivers. Usually, such filters are mounted on the radiation shield of the cryostat to reduce the heat load to the 4 K stage. Insufficient filtering may cause the temperature of the mixing element in a receiver to be excessively warm, leading to degradation in sensitivity. These filters should be effective in blocking the room temperature IR radiation from outside the cryostat, yet should be transparent across the desired signal frequency band. In the Terahertz frequency range, which is close to the infrared, it is difficult to find an inexpensive low- loss material that can provide the required IR blocking capacity. We present transmission measurements, made using a Fourier Transform Spectrometer (FTS), of a number of potential infrared filters between 0.4 and 1.6 THz. The filters tested include the widely-used, Teflon-based, Zitex-A and Zitex-G films, alkali halide based infrared filter, and crystalline quartz coated with Parylene, and polyethylene films.
|
|
|
Cherednichenko, S., Kollberg, E., Angelov, I., Drakinskiy, V., Berg, T., & Merkel, H. (2005). Effect of the direct detection effect on the HEB receiver sensitivity calibration. In Proc. 16th Int. Symp. Space Terahertz Technol. (pp. 235–239). Göteborg, Sweden.
Abstract: We analyze the scale of the HEB receiver sensitivity calibration error caused by the so called “direct detection effect”. The effect comes from changing of the HEB parameters when whey face the calibration loads of different temperatures. We found that for HIFI Band 6 mixers (Herschel Space Observatory) the noise temperature error is of the order of 8% for 300K/77K loads (lab receiver) and 2.5% for 100K/10K loads (in HIFI). Using different approach we also predict that with an isolator between the mixer and the low noise amplifiers the error can be much smaller.
|
|
|
Loudkov, D., Tong, C. - Y. E., Blundell, R., Kaurova, N., Grishina, E., Voronov, B., et al. (2005). An investigation of the performance of the waveguide superconducting HEB mixer at different RF embedding impedances. In Proc. 16th Int. Symp. Space Terahertz Technol. (pp. 226–229).
Abstract: We have conducted an investigation of the performance of superconducting hot-electron bolometric (HEB) mixer at 800 GHz as a function of the embedding impedance of the waveguide embedding circuit. Using a single half-height mixer block, we have developed three different mixer chip configurations, offering nominal embedding resistances of 70, 35, and 15 Ohms. Both the High Frequency Structure Simulator (HFSS) software and scaled model impedance measurements were employed in the design process. Two batches of HEB mixers were fabricated to these designs using 3-4 nm thick NbN thin film. The mixers were characterized through receiver noise temperature measurements and Fourier Transform Spectrometer (FTS) scans. Briefly, a minimum receiver noise temperature of 440 K was measured at a local oscillator frequency 850 GHz for a mixer of normal state resistance 62 Ohms incorporated into a circuit offering a nominal embedding impedance of 70 Ohms. We conclude from our data that, for low noise operation, the normal state resistance of the HEB mixer element should be close to that of the embedding impedance of the mixer mount.
|
|
|
Yang, Z. Q., Hajenius, M., Baselmans, J. J. A., Gao, J. R., Klapwijk, T. M., Voronov, B., et al. (2005). Improved sensitivity of NbN hot electron bolometer mixers by vacuum baking. In Proc. 16th Int. Symp. Space Terahertz Technol. (pp. 222–225).
Abstract: We find that the sensitivity of heterodyne receivers based on superconducting hot-electron bolometer (HEB) in- creases by 25 − 30% after baking at 85 o C and in a high vacuum. The devices studied are twin-slot antenna coupled HEB mixers with a small NbN bridge of 1×0.15 μm 2 . The mixer noise temperature, gain, and resistance versus temperature curve of a HEB before and after baking are compared and analyzed. We show that baking reduces the intrinsic noise of the mixer by 37 % and makes the superconducting transition of the bridge and the contacts sharper. We argue that the reduction of the noise is due to the improvement of the transparency of the contact/film interface. The lowest receiver noise temperature of 700 K is measured at a local oscillator frequency of 1.63 THz and a bath temperature of 4.3 K.
|
|