PT Unknown AU Seliverstov, SV Rusova, AA Kaurova, NS Voronov, BM Goltsman, GN TI Attojoule energy resolution of direct detector based on hot electron bolometer SE J. Phys.: Conf. Ser. PY 2016 BP 012165 (1 to 5) VL 741 DI 10.1088/1742-6596/741/1/012165 DE NbN HEB detector AB We characterize superconducting antenna-coupled NbN hot-electron bolometer (HEB) for direct detection of THz radiation operating at a temperature of 9.0 K. At signal frequency of 2.5 THz, the measured value of the optical noise equivalent power is 2.0×10-13 W-Hz-0.5. The estimated value of the energy resolution is about 1.5 aJ. This value was confirmed in the experiment with pulsed 1.55-μm laser employed as a radiation source. The directly measured detector energy resolution is 2 aJ. The obtained risetime of pulses from the detector is 130 ps. This value was determined by the properties of the RF line. These characteristics make our detector a device-of-choice for a number of practical applications associated with detection of short THz pulses. ER