%0 Journal Article %T A low temperature bolometer heterodyne receiver for Millimeter wave astronomy %A Phillips, T. G. %A Jefferts, K. B. %J Rev. Sci. Instrum. %D 1973 %V 44 %N 8 %F Phillips+Jefferts1973 %O Recommended by Klapwijk %O exported from refbase (https://db.rplab.ru/refbase/show.php?record=927), last updated on Tue, 24 Jun 2014 17:17:23 -0500 %X Liquid helium cooled InSb hot electronbolometers are used in a balanced mixer configuration as detectors for an imagelessmicrowave receiver. The system is designed for mounting at the prime focus of the National Radio Astronomy Observatory (NRAO) 11 m antenna at Kitt Peak, Arizona, and is suitable for the study of rotational line spectra of interstellar gas molecules. Currently the operating frequency is in the 90–140 GHz band where the double sideband system noise temperature is 250 K. %K InSb HEB mixer %R 10.1063/1.1686288 %U https://doi.org/10.1063/1.1686288 %P 1009-1014